background image

Short instructions for use 1 Controls

6

1.4 Groups of keys
Group of keys for the dental chair

The keys of the assistant unit each have two functions and show two symbols.

Assistant element key

Dentist element key

Name
"Chair up" key

"AP 0" key

(automatic position 0)

"Chair down" key

"SP" key

(rinsing position)

"LP" key

(last position)

"AP" key

(activate automatic position)

"Backrest down" key

"AP 1" key

(automatic position 1)

"Backrest up" key

"AP 2" key

(automatic position 2)

"Collapsed position" key

Group of keys for illumination

Key

Name

Control element

"Operating light" key

Assistant element

"X-ray image viewer" key

Dentist element

Group of keys for hygiene

Key

Name

Control element

"Tumbler filler" key

Dentist element and

assistant element

Summary of Contents for KaVo Primus 1058 Life TM

Page 1: ...Short instructions for use Primus 1058 Life...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...r time 15 6 Setting the handpieces 16 6 1 Adjusting functions 16 6 2 Turbine 18 6 3 INTRA LUX Motor KL 703 LED and COMFORTdrive 18 6 4 PiezoLED 18 6 5 Multifunctional handpiece 19 7 Using handpieces 2...

Page 4: ...de all information and safety notes that must be observed to ensure intended use and safe operation of the product The user must ensure that the unit works properly and is in satisfactory condition be...

Page 5: ...al chair E Group of keys for illumination F Group of keys for hygiene 1 3 Assistant element A Group of keys for hygiene B Group of keys for illumination C Group of keys for the timer D Group of keys f...

Page 6: ...n 0 Chair down key SP key rinsing position LP key last position AP key activate automatic position Backrest down key AP 1 key automatic position 1 Backrest up key AP 2 key automatic position 2 Collaps...

Page 7: ...ptional HYDROclean key Assistant element Group of keys for the handpieces timer Key Name Control element Preselected spray button Dentist element Direction of motor rotation but ton Dentist element Ti...

Page 8: ...ir motion function Blown air AP footswitch Moves the dental chair into automatic position Sets the preset blown air chip blower Cross switch Counterclock wise motor rotation Changes the position of th...

Page 9: ...Page 8 2 1 5 Foot control Page 8 2 3 Automatic positioning of patient chair Recalling automatic position through the dentist element Briefly press the desired button Chair automatically moves to the...

Page 10: ...until a beep is issued LED of the pressed button lights up The chair position is saved Recalling an automatic position through the foot control Press the SP foot operated button or Press the LP foot o...

Page 11: ...Set time of day 3 Date Set date 4 Time display mode Setting the time of day display mode Time of day only Time of day without seconds 5 Language Set menu language Deutsch English Italiano Fran ais 6 L...

Page 12: ...e Save key S1 to exit from the user menu The changed settings are saved 3 2 Selecting and setting an option Press the Decrease value or the Increase value keys to select the desired pa rameter or valu...

Page 13: ...ction keys Display Switching between levels Available with MEMOspeed licence only Level switching is deactivated in the basic state The level switching symbol displays the current dentist Note The dev...

Page 14: ...for the current den tist Note Level switching is deactivated using the same key combination as activation Briefly press Preselect level key to select the level Selection of dentist Level switching bl...

Page 15: ...imer Requirement Standby menu is selected To set a timer time e g Timer 1 press the Timer key until you hear a beep The display switches to the settings menu for the timer time Press the Decrease valu...

Page 16: ...level switching in level 1 3 Saving the values Press the Save key to save the values You can save after setting each value or after setting all values Saving is acknowledged by a beep This closes the...

Page 17: ...ting the direction of motor rotation Note The direction of motor rotation can only be changed when the motor is at rest Requirement Settings menu for motor is selected Press the Direction of motor rot...

Page 18: ...s for Use Press the Mode S4 key for 4 seconds to select the operating mode Modes P1 P2 P3 E are available for selection Symbol Feature Mode P1 Mode E 6 2 Turbine The following settings can be changed...

Page 19: ...can be changed Output intensity Operating mode P1 P2 P3 E Cooling status no cooling spray water cooling Settings menu for minimal maximal intensity 6 5 Multifunctional handpiece The following settings...

Page 20: ...ds to the minimum speed intensity The right stop corresponds to the maximum speed intensity Preselect counterclockwise motor rotation Take motor off the holder The handpiece is active Slide the cross...

Page 21: ...rating light On and Off Note Activate the KaVoLUX 540 LED operating light using the Operating light key on the assistant element Only then the operating light can be operated by means of the sensor an...

Page 22: ...the Mode colour temperature button The treatment light switches from COMPOsave mode to dimmed normal light Press the Mode colour temperature button once again The treatment light switches back to COMP...

Page 23: ...ching ring to the left until it snaps into place Treatment light can be rotated 45 to the left or 45 to the right Turn the switching ring to the right and it will spring back to its original position...

Page 24: ...cam U ERGOcam One 3 Next image video To be able to communicate efficiently with the patient individual im ages can be selected and displayed directly This uses a rolling sys tem that advances from lef...

Page 25: ...selected images from the clipboard are displayed When the patient is logged off the corresponding PC the clipboard is deleted 9 1 Using the patient communication menu with the foot control The patient...

Page 26: ...in the clipboard Note If no patient is selected the images stay in the Swap Tray and are not saved per manently As soon as a patient is selected these temporary data in the Swap Dry ing are deleted W...

Page 27: ...than 4 seconds to start the programming mode Set the period of time in 200 ms increments Minimum 0 4 s An acoustic signal is heard every second Release the button to save the selected value 10 2 Rinsi...

Page 28: ...annula with a slight twisting motion Insert the triple function or multifunctional handpieces without the cannula into the clamping de vice of the germ reduction attachment Make sure that the W button...

Page 29: ...insing programme The rinsing programme can be stopped at any time by pressing the intensive germ reduction key If the device is turned off and then on again the rinsing program also stops 10 3 Intensi...

Page 30: ...L 6 concentration Requirement intensive germ reduction is in program step 4 You hear a beep for 4 seconds Turn the device off When the device is switched on after 30 minutes of dwell time the remainin...

Page 31: ...n hoses in the holder 10 5 HYDROclean function The following processes are triggered by the HYDROclean function The suction lines from the filter housing to the separator system are rinsed automatical...

Page 32: ...sued every 10 seconds Remove the tumbler filler and tumbler holder Top off the Oxygenal max 1 litre Note Do not fill through the centre opening tumbler drain Quit replenishing immediately if you hear...

Page 33: ...d patient chair Item No Safety switch off actuated LED on assistant element LED on dentist element Patient unit swung over the patient chair Assistant element Backrest Bracket on the foot control Kick...

Page 34: ...ounds The amalgam separator CAS1 is almost full 95 Exchange the amalgam container The CAS1 amalgam separator is defective Also refer to Instructions for use for the CAS 1 or Call a Service technician...

Page 35: ...emedy Display shows ID 33 No CAN node and or in ternal communication is faulty Call a service technician Display shows ID 64 Water is shut off Turn water on Water works leaks strongly Water works malf...

Page 36: ...0 Dekaseptol empty Replenish Dekaseptol See also 2 Servicing instructions Display shows ID 72 Dekaseptol bottle Insert the DEKASEPTOL bottle See also 2 Care instructions Display shows ID XX This error...

Page 37: ......

Page 38: ...1 011 8577 Fk 20160324 3 en...