Short instructions for use 10 Hygiene
31
Application of DEKASEPTOL Gel
▶ Aspirate one tumbler of cold water with each suction hose.
▶ With the suction handpiece attached or by hand, press down the Dekaseptol dosing facility once
each for the saliva ejector hose and the spray mist ejector.
▶ Aspirate DEKASEPTOL Gel (Mat. no. 10007205) from the dosing facility.
▶ Aspirate another tumbler of cold water with each suction hose.
▶ Then aspirate another dose of Dekaseptol and allow it to act.
▶ Place the suction hoses in the holder.
10.5 HYDROclean function
The following processes are triggered by the HYDROclean function:
• The suction lines (from the filter housing to the separator system) are rinsed automatically for 150
seconds with pure water by activating the selective valves and the suction machine.
▪ The bowl rinse is turned on and off several times.
Requirement
Suction hoses remain hanging in the holder.
▶ Press the "HYDROclean" button.
ð
The HYDROclean function is in the preparatory stage.
▶ Press the "HYDROclean" button.
ð
This starts-up the HYDROclean function.
The HYDROclean function is completed when the top LED goes dark.
The process can be discontinued at any time by pressing the "HYDROclean" button.