CAUTION

Contact corrosion from moisture.

Damage to product

▶ Immediately remove the product from the steam steriliser after the

sterilisation cycle!

The medical device is temperature resistant up to 138℃ (280.4°F).

KaVo recommends for example

- STERIclave B 2200/ 2200P by KaVo

- Citomat/ K-series by Getinge

 

Preparation methods according to ISO 17664

47

Summary of Contents for INTRAmatic contra-angle 20 ES

Page 1: ...Instructions for use INTRAmatic contra angle 20 ES REF 1 003 4830...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...ditions 6 2 Safety 8 2 1 Description of safety instructions 8 2 2 Purpose Intended use 10 2 3 Safety instructions 12 3 Product description 15 3 1 Technical Specification 15 3 2 Transportation and stor...

Page 4: ...nder 28 6 Preparation methods according to ISO 17664 30 6 1 Preparations at the site of use 30 6 2 Cleaning 31 6 2 1 Cleaning Manual cleaning external 31 6 2 2 Cleaning Automated external cleaning 32...

Page 5: ...are products and systems Servicing 41 6 5 1 Care products and systems Servicing Care with KaVo Spray 42 6 5 2 Care products and systems Servicing Care with the KaVo SPRAYrotor 43 6 5 3 Care products a...

Page 6: ...ing this KaVo quality product Following the in structions below will allow you to work smoothly economically and safely Copyright by KaVo Dental GmbH Symbols Refer to chapter Safety Warning Important...

Page 7: ...unaut Europ enne A product with this mark meets the requirements of the applicable EC directive Action request Target group This document is intended for dentists and their assistants The section on s...

Page 8: ...air Other claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory l...

Page 9: ...onsequences that can be attributed to interventions on or changes made to the product by the customer or a third party not authorised by KaVo are excluded from the warranty Service warranty claims wil...

Page 10: ...uctions Warning symbol Structure DANGER The introduction describes the type and source of the danger This section portrays the possible consequences of non observance The optional step covers necessar...

Page 11: ...hree levels of danger will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild or moderate physical harm WARNING WARNING ind...

Page 12: ...or dental treatment Any other type of use or alteration to the product is impermissible and can be hazardous The medical device is intended for the following uses Removal of carious material cavity pr...

Page 13: ...able health and safety regulations the applicable accident prevention regulations these instructions for use According to these regulations the user is required to Only use equipment that is operating...

Page 14: ...CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Reduced production time The instrument must be cleaned serviced and stored dry if it has not been used for...

Page 15: ...cutter or grinder cannot be firmly held Stop work and seek service support CAUTION Hazard from improperly putting away instruments Injury and infection caused by chucked cutters or grinders Damage to...

Page 16: ...st be set up according to the methods described in the KaVo instructions for use and the care prod ucts and methods described therein must be used KaVo recommends specifying a service interval at the...

Page 17: ...3 Product description INTRAmatic contra angle handpiece 20 ES Mat no 1 003 4830 3 1 Technical Specification Drive speed identification 1 blue ring Product description 15...

Page 18: ...With pushbutton chuck Short handpiece cutters or grinders can be used The contra angle handpiece can be mounted on all INTRAmatic LUX mo tors and motors with a connection in accordance with ISO 3964 D...

Page 19: ...l device after it has been stored re frigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F Tempe...

Page 20: ...Air pressure 700 hPa to 1 060 hPa 10 psi to 15 psi Protect from moisture Product description 18...

Page 21: ...rovider and patient Before first use and after each use sterilise the medical device 4 1 Check the amount of water CAUTION Overheating of the tooth due to lack of cooling water Thermal damage to the p...

Page 22: ...ent spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 09...

Page 23: ...n 5 1 Attach the medical device CAUTION Removing and attaching the handpiece while the drive motor is rotating Damage to the catch Never attach or remove the handpiece while the device is rotating Ope...

Page 24: ...e A medical device that is not properly locked in place can release from the motor coupling and fall off Carefully pull on it before each treatment to ensure that the medical device is securely locked...

Page 25: ...ing and lock it into place Before each treatment pull on the medical device to see if it is securely seated in the motor coupling 5 2 Remove the medical device Remove the medical device from the LUX m...

Page 26: ...mond grinders Note Only use carbide cutters or diamond drill bits that correspond to ISO 1797 1 type 1 are made of steel or hard metal and meet the following criteria Shaft diameter 2 334 to 2 35 mm O...

Page 27: ...nstructions for use and use the cutter or grinder prop erly Only use cutters or grinders that do not deviate from the indicated data CAUTION Injury from using worn cutters or grinders Cutters or grind...

Page 28: ...ard from defective chucking system The cutter or grinder can fall out and cause injury Pull on the cutter or grinder to check if the chucking system is okay and the cutter or grinder is securely held...

Page 29: ...Insert the cutter or grinder into the segment of the head drive by twisting the tool slightly and push to the stop Check that the tool is seated by pulling on it Operation 27...

Page 30: ...Do not touch the cutter or grinder when it is rotating Never press the pushbutton while the cutter or grinder is rotating Do not touch soft tissue with the head tip that can heat and burn the soft tis...

Page 31: ...After the cutter or grinder has stopped rotating press the push button with your thumb and pull out the cutter or grinder Operation 29...

Page 32: ...medical devices Observe suitable personal protective measures Remove residual cement composite or blood at the site of use Dry the medical device to prepare it for transportation Do not place it in a...

Page 33: ...fects in the product Only clean in a thermodisinfector or manually 6 2 1 Cleaning Manual cleaning external Required accessories Tap water 30 C 5 C 86 F 10 F or a 60 to 70 alcohol solution Brush such a...

Page 34: ...ds thermodisinfectors in accordance with ISO 15883 1 e g Miele G 7781 G 7881 Validation was performed with the program VARIO TD the cleaner ne odisher mediclean the neutraliser neodisher Z and rinse n...

Page 35: ...internal For a medical device with KaVo CLEANspray and KaVo DRYspray Cover the medical device with the KaVo CLEANpac bag and place it on the corresponding care adapter Press the spray button three ti...

Page 36: ...Luxem bourg Belgium Netherlands United Kingdom Denmark Sweden Fin land and Norway In other countries only automated interior cleaning with thermodisinfectors in accordance with ISO 15883 1 6 2 4 Clea...

Page 37: ...refers to the material compatibility with KaVo products The program settings and cleansers and disinfectants that must be used can be found in the instructions for use of the thermodisinfector Direct...

Page 38: ...ion CAUTION Malfunctioning from using a disinfectant bath or chlorine containing dis infectant Defects in the product Only disinfect in a thermodisinfector or manually Preparation methods according to...

Page 39: ...ed on material compatibility The microbiological efficacy must be ensured by the disinfectant manufac turer Microcide AF by Sch lke Mayr liquid or cloths FD 322 by D rr CaviCide by Metrex Required too...

Page 40: ...ectant 6 3 2 Disinfection Manual disinfection internal To effectively set up the inside of the machine must be cleaned automat ically in a cleaning and disinfection unit in accordance with ISO 15883 1...

Page 41: ...r and only refers to the material compatibility with KaVo products The program settings and cleansers and disinfectants that must be used can be found in the instructions for use of the thermodisinfec...

Page 42: ...pressed air until no water drops are visible Machine drying The drying procedure is normally part of the cleaning program of the ther modisinfector Note Please observe the instructions for use of the...

Page 43: ...e medical device Injury hazard from sharp and or pointed cutters or grinders Remove the cutter or grinder CAUTION Premature weary and malfunctions from improper servicing and care Reduced production t...

Page 44: ...per use on our products 6 5 1 Care products and systems Servicing Care with KaVo Spray KaVo recommends servicing the project twice daily at noon and in the evening after hours after each time the mach...

Page 45: ...nds servicing the project twice daily at noon and in the evening after hours after each time the machine is cleaned and before each sterilisation Place the product on the appropriate coupling of the K...

Page 46: ...nsion pressure for effective cleaning and care KaVo recommends servicing the project twice daily at noon and in the evening after hours after each time the machine is cleaned and before each sterilisa...

Page 47: ...tretched The quality and use of the sterilisation packaging must satisfy applicable standards and be suitable for the sterilisation procedure Individually weld the medical device in the sterilised ite...

Page 48: ...oclave EN 13060 ISO 17665 1 CAUTION Premature weary and malfunctions from improper servicing and care Reduced production time Before each sterilisation cycle treat the medical device with KaVo care pr...

Page 49: ...remove the product from the steam steriliser after the sterilisation cycle The medical device is temperature resistant up to 138 280 4 F KaVo recommends for example STERIclave B 2200 2200P by KaVo Cit...

Page 50: ...Autoclave using the gravitational method sterilise for at least 10 min at 134 C 1 C 273 F 1 8 F Autoclave using the gravitational for at least 60 min at 121 C 1 C 250 F 1 8 F Follow the manufacturer...

Page 51: ...Note Observe the expiration date of the sterilized item Preparation methods according to ISO 17664 49...

Page 52: ...pliers Material summary Mat no Instrument stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0931 Silicone hose 150 mm 0 593 0252 KaVo CLEANspray...

Page 53: ...Material summary Mat no KaVo Spray 2112 A 0 411 9640 ROTAspray 2142 A 0 411 7520 QUATTROcare plus Spray 2108 P 1 005 4525 Tools 51...

Page 54: ......

Page 55: ......

Page 56: ...1 003 7238 kb 20090903 02 en...

Reviews: