Care with KaVo SPRAYrotor
Removing the bur.
Attach medical product to the appropriate
coupling of the SPRAYrotor and cover with a
Cleanpac bag.
Carry out care of medical product twice daily,
at midday and in the evening after the practice
closes, and after each preparation in a
machine and before each sterilization.
Use according to the instructions for use for
the KaVo SPRAYrotor.
Care with QUATTROcare
Care & maintenance unit utilising oil and
detergent under pressure, for a good cleaning
and lubricating result.
Removing the bur.
Carry out care of medical product twice daily,
at midday and in the evening after the practice
closes, and after each preparation in a machine
and before each sterilization.
Use according to the instructions for use for
the KaVo QUATTROcare.
Sterilisation in steam steriliser
Before each sterilisation cycle, you have to prepare the medical device
using KaVo care products.
The medical device can be sterilised in autoclaves, e.g. LIC up to a temperature of
135 °C.
Packaging:
Package each instrument individually in STERIclave bags. The STERIclave bag must be
large enough to cover the instrument and prevent the packaging from becoming
electrically energised.
Sterilisation:
Autoclave with triple pre-vacuum, 3 minutes minimum at 134 °C ± 1
Autoclave with the gravitation process, 10 minutes minimum at 134 °C ± 1
Autoclave with the gravitation process, 60 minutes minimum at 121 °C ± 1
Use in accordance with manufacturer's user instructions.
Immediately after every sterilisation cycle, the medical device must be removed
from the steriliser to prevent the risk of contact corrosion.
Accessories
Are supplied by your Dental Depot.
Description
Mat.-No.
Instrument stand 2151
0.411.9501
Cellulose pads, 100 pcs.
0.411.9862
Cleanpac, 10 pcs.
0.411.9691
Bur stop
0.524.0892
Hook
0.410.1633
KAVOspray 2112 A
0.411.9640
ROTAspray 2142 A
0.411.7520
QUATTROcare Spray 2108 A
0.411.7720
LIFEtime Spray 2118 A
0.411.9700
Further advice for use and safety
When the INTRAmatic handpiece will not be used for an extended period of
time, it should be cleaned and lubricated according to the instructions and
stored in a dry place.
If damaged, irregular noise when running, too much vibration, too much drill
impact or if drill is not held, do not carry on.
To avoid further damage, please contact the KaVo Customer Service Center at
Warthausen, or a KaVo authorised office or your Dental Depot.