▪ Autoclave using the gravity method:
- at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▶ Use according to the manufacturer's Instructions for Use.
7.7 Storage
Prepared products must be stored, protected from germs (as far as possi‐
ble) and dust, in a dry, dark, cool room.
Note
Comply with the expiry date of the sterilised items.
7 Reprocessing methods according to ISO 17664