![KaVo INTRAflex LUX 2 program 2307 LN 0.535.6190 Instructions For Use Manual Download Page 34](http://html1.mh-extra.com/html/kavo/intraflex-lux-2-program-2307-ln-0-535-6190/intraflex-lux-2-program-2307-ln-0-535-6190_instructions-for-use-manual_658172034.webp)
▶ Insert the test gauge (Mat. no. 04118731) between the hose and the
MULTIflex (LUX) / MULTIflex LED coupling and check the following
pressures:
ð
- Drive air: 2.5 to 3.0 bar (29 to 44 psi), recommended: 3.0 (44 bar psi)
ð
Return air: < 1 bar (15 psi)
ð
- Water: 1.0 to 2.0 bar (15 to 29 psi)
ð
- Spray air: 1.5 to 2.5 bar (22 to 36 psi)
4 Start up and shut down