Autoclave with a triple pre-vacuum for at least 4 minutes at 134 °C ± 1 °C (273 °F ± 1.8 °F)
Autoclave with gravitation process for at least 10 minutes at 134 °C ± 1 °C (273 °F ± 1.8 °F)
Autoclave with gravitation process for at least 60 minutes at 121°C ± 1°C (250°F ± 1.8°F)
Comply with the manufacturer's instructions for use with regard to the application range.
6.8
Storage
Prepared products must be stored, protected from germs (as far as possible) and dust, in a dry,
dark, cool room.
Reprocessing methods according to ISO 17664