Note
Follow the instructions for use of the disinfectant.
6.3.2
Disinfection: Manual internal disinfection
For the effective re-preparation, the inside of the device must be cleaned automatically in a cleaning
and disinfection unit in accordance with ISO 15883-1.
The interior of this product is not designed for manual disinfection.
Reprocessing methods according to ISO 17664
55