Tools required:
▪ Cloths for wiping down the medical device.
▶
Spray the disinfectant on a cloth, then thoroughly wipe down the med‐
ical device and leave the disinfectant to soak in according to the in‐
structions from the disinfectant manufacturer.
Note
Observe the instructions for use for the disinfectant.
6.5.2 Disinfection: Manual disinfection - internal
For the effective re-preparation, the inside of the device must be cleaned
automatically in a cleaning and disinfection unit in accordance with ISO
15883-1.
The interior of this product is not designed for manual disinfection.
Preparation methods according to ISO 17664
37