5.3 Insert the milling cutters or diamond grinders
Note
Only use carbide cutters or diamond grinders that correspond to ISO
1797-1 type 3, are made of steel or hard metal and meet the following
criteria:
- shaft diameter: 2.334 to 2.350 mm without drill stop
- Overall length: max. 70 mm
- Shaft clamping length: min. 56 mm
- Cutting diameter: max. 3 mm
Operation
28
Summary of Contents for INTRA 3610 N3
Page 1: ...Instructions for use INTRA surgical handpiece 3610 N3 REF 0 524 5620...
Page 21: ...Air pressure 700 hPa 10 psi to 1060 hPa 15 psi Protect from moisture Product description 21...
Page 24: ...Do not use a different coolant First use 24...
Page 34: ...Remove the drill stop using the accompanying hook Operation 34...