Technician's instructions KaVo In eXam 3510
1 Product description
|
1.1 Description of electronics
▶ Turn on the In eXam
▶ Turn on the In eXam
▶ Note the no-load voltage: U
o
= 230 V (+6%/-10%)
▶ Note the no-load voltage: U
o
= 230 V (+6%/-10%)
▶ Set parameters: Film type 9, 60kV, 7mA, occlusal mode
▶ Observe radiation protection measures
▶ Trigger the exposure
▶ Note the load voltage (U
load
) during exposure
▶ Calculate the voltage fluctuation (U
o
– U
load
< 9V [~4%])
▶ When the voltage fluctuation > 9V: Check the power cable
+ 36 V
7/32