![KaVo GENTLEpower LUX handpiece 10 LP Instructions For Use Manual Download Page 44](http://html1.mh-extra.com/html/kavo/gentlepower-lux-handpiece-10-lp/gentlepower-lux-handpiece-10-lp_instructions-for-use-manual_3841170044.webp)
▶
For program settings as well as cleansers and disinfectants to be
used, please refer to the Instructions for Use of the thermodisinfector
(complying with max. pH value of 10).
▶
In order to prevent negative effects on the medical device, make
sure that the interior and the exterior of the medical device are dry
after completion of the cycle, and then grease it immediately with
servicing agents from the KaVo care system.
6.4 Drying
Manual Drying
▶
Blow off the outside and inside with compressed air until water drops
are no longer visible.
Preparation methods according to ISO 17664
42
Summary of Contents for GENTLEpower LUX handpiece 10 LP
Page 1: ...Instructions for use GENTLEpower LUX handpiece 10 LP 1 001 5660...
Page 6: ...7 Tools and consumables 61 8 Warranty terms and conditions 63 Table of contents 4...
Page 15: ...3 Product description GENTLEpower LUX Handpiece 10 LP Mat no 1 001 5660 Product description 13...
Page 20: ...Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 18...
Page 23: ...3 Start up and shut down 21...
Page 67: ......
Page 68: ...1 001 9319 kb 20130802 06 en...