Spray the disinfectant on a cloth, then wipe down the medical device
and allow the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
6.3.2 Disinfection: Manual disinfection - internal
The efficacy of manual internal disinfection must be demonstrated by the
manufacturer of the disinfection agent. With KaVo products, use only dis‐
infection agents that have been released by KaVo with respect to the
compatibility of materials
(e.g. WL-cid / made by ALPRO).
▶
Immediately after internal disinfection, lubricate the KaVo medical
device immediately with care agents from the KaVo care system.
Preparation methods according to ISO 17664
40