Instructions for use FLEXspace Typ 5200, 5200-U, 5210, 5210-U
3 Description of the product | 3.2 Technical Data
Gas
Max. gas pressure at customer
premises
50 mbar
Permissible gases:
In accordance with DVGW Work Sheet G260/I
and ANSI Z21.1/CSA 1.1 pp.
Max. gas consumption
Natural Gas
0.1 m
3
/h
Liquid gas
0.075 kg/h
City gas
0.15 m
3
/h
Nominal voltages, nominal frequency
Type 5200-UL, 5210 UL
100-120 V 50/60 Hz
Type 5200, 5210
230 V 50/60 Hz
Line voltage fluctuations
±10 %
Power consumption
max. 12 A
UL version: Power rating of the
socket strip
max.12 A
CAUTION
Overloading of electrical cables
Electrical shock
▶
The total power of the connected electrical loads must not exceed 12 A, but
can be distributed over the number of installed sockets.
▶
This limitation must also be ensured if 2 workstations are connected via one
fuse protection (max. 12 A).
▶
The same must be ensured if 3 workstations are connected via one power
supply cord.
Leakage current
max 3.5 mA
Fuse protection in accordance with
national requirements, but
max. 16 A
Overvoltage category
II
Degree of soiling
2
Class of protection
I
13 / 38