Page 1: ...Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 Always be on the safe side ...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Page 3: ...Safety instructions 2 2 Product description 3 2 1 Technical data 3 3 Exchanging the spare turbine 4 4 First use 6 5 Function test 7 6 Tools and consumables 8 Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 Table of contents 1 8 ...

Page 4: ...pment for the proper purpose to protect him or herself the patient and third parties from hazards to avoid contamination by the product 1 2 Safety instructions CAUTION Injury or damage due to wear Irregular running noise significant vibration overheating imbalance or insuffi cient grip Stop work and seek service support Note KaVo accepts no liability for damage arising from repairs that have been ...

Page 5: ...nt turbine EXPERTtorque mini with and without wrench Mat no 1 007 9458 and Mat no 1 007 9457 consisting of 1 x replacement turbine assembled 1 x wrench optional 2 x O rings 1 x Spring washer larger outer diameter shiny 1 x Spring washer smaller outer diameter coloured brown Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 2 Product description 2 1 Technical data 3 8...

Page 6: ...e wrench toward the cover and unscrew the cover in counterclockwise direction Take out the replacement turbine Use tweezers to push out and take out the replacement turbine from the front Remove the O rings Remove the O rings from the head housing and cover with tweezers Remove the spring washers Remove the spring washers form the head casing and cover using tweezers Clean Before installation of t...

Page 7: ...ct position of the O rings Functional problem Make sure that the O rings are positioned correctly in the recesses Replace both O rings Spray O rings with KaVo Spray and insert them in the head housing and cov er Insert the replacement turbine DANGER Not maintaining a torque of 100 Ncm when tightening the cover Danger of injury the cover can release and fall into the patient s mouth with the replac...

Page 8: ...ar Ensure that the cooling air is dry clean and uncontaminated in accordance with EN ISO 7494 2 Spray the KaVo recommended care product through the medical product and attach the medical product to the MULTIflex coupling Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 4 First use 6 8 ...

Page 9: ...tart the medical device Perform a test run under axial and radial load Tighten the cover again applying a torque of 100 Ncm 6 Ncm Note If the unit runs poorly or heats up in untypical manner please return the entire turbine contra angle handpiece to KaVo for testing Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 5 Function test 7 8 ...

Page 10: ...uter diameter shiny 1 003 5900 Spring washer smaller outer diameter coloured brown 1 003 7576 Material summary Mat No Instrument stand 2151 0 411 9501 Insert for medical device 0 411 9902 Cleanpac 0 411 9691 KaVo Torque wrench 1 4 100 Ncm 1 002 4578 Instructions for use Spare cartridge for EXPERTtorque mini 1 007 9458 1 007 9457 6 Tools and consumables 8 8 ...

Page 11: ......