7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual Reprocessing
20 / 28
7.2 Manual Reprocessing
WARNING
Incomplete disinfection.
Patients, users or third parties could be infected by contaminated medical
devices.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
WARNING
Sharp dental bur in the medical device.
Injury hazard from sharp and/or pointed dental bur.
4
Remove the dental bur.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
7.2.1 Manual pre-cleaning
4
Rinse the instrument for 20 seconds with process water on the treatment
centre.
4
Remove instrument from the treatment centre and rinse for 20 seconds un-
der running tap water.
4
Attach fitting adapter to the CLEANspray can.
4
Attach instrument with the drive side to the adapter.
4
Rinse through the instrument for at least 3 x 2 seconds (solution must flow
out visibly).
4
Take the instrument from the adapter and set it down.
4
Spray the surface of the instrument with CLEANspray until the entire sur-
face is wet.
4
Allow the cleaning agent to act for 5 minutes.
4
Rinse the product with tap water for at least 10 seconds.
4
Remove visible contaminants with a soft, PVC brush.
7.2.2 Manual internal and external cleaning and internal
and external disinfection
Manual cleaning
Accessories required:
▪ Tap water (drinking water)
▪ Brush, e.g. medium-hard toothbrush
4
Brush off under flowing tap water at least 10 seconds.