Instructions for use EXPERTmatic E10 C – 1.007.5560
3 Description of the product | 3.1 Intended use
10 / 28
3 Description of the product
EXPERTmatic Handpiece E10 C (
Mat. no. 1.007.5560
)
3.1 Intended use
Indications for use:
This medical device is:
▪ Intended for dental treatment only. All other types of use of or modifications
to the product are not permitted and can be hazardous.
▪ The medical device is intended for the following applications:
– Removal of carious material
– Cavity and crown preparations
– Removal of fillings
– Processing of tooth and restoration surfaces
▪ A medical device according to relevant national statutory regulations.
Proper use:
According to these regulations, this product may only be used for the described
application by a properly trained user. You need to comply with the following:
▪ the applicable health and safety regulations
▪ the applicable accident prevention regulations
▪ these Instructions for use
According to these regulations, the user is required:
▪ exclusive use of equipment that is operating correctly.
▪ compliance with the specified intended use.
▪ Protect themselves, patients and third parties from danger.
▪ Contamination by the product must be avoided.
3.2 Technical specifications E10 C (without light)
Drive speed
max. 40,000 rpm
Speed transmission
1 : 1
Identification
1 blue ring
Spray water pressure
0.8 to 2.0 bar (12 to 29 psi)
Spray air pressure
1.0 to 2.0 bar (15 to 29 psi)
Spray air flow rate
min. 1.5 Nl/min (at 2 bar)
Spray water flow rate
min. 50 ml/min (at 2 bar)
Cooling air flow
5.5 to 9.5 Nl/min