Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.15 Using the SL600 surgical motor (optional accessory)
124 / 178
Mode
Note
30 seconds operating time/9 minutes pause is the potential load threshold of
the motor (full load at maximum speed).
4.15.2 Connecting and operating the pump for
physiological saline
The surgical motor set comes with the kit "pump for physiological saline solu-
tion".
See also:
2
Assembly instructions for saline solution kit
4.15.3 Connecting the SL 600 surgery motor
Note
The surgical mode can only be accessed when the surgical motor is connec-
ted to the surgical connection of the dentist element.
Note
The delivered parts are not sterile (except for the coolant hose). Before the
first treatment of a patient, the surgical motor, motor cable, and the hand-
piece tray need to be reprocessed.
1
2
3
4
①
Motor coupling
②
Surgical motor
③
Handpiece tray
④
Plug of motor cable
▶
Plug the surgical motor
②
into the motor coupling
①
and secure it with a
union nut. Please note the separate instructions for use of the motor.
▶
Place the surgical motor on the handpiece tray
③
.
▶
Insert the plug of the motor cable
④
into the connector on the device, align
the marker points, and insert the plug until it snaps into place.