Short instructions for use 6 Hygiene
35
▶
During the active phase, do not place hoses in holders and, if applicable, do not
switch off the treatment centre.
▶
Wait for the end of the active phase or switch on the treatment centre.
ð
A progress bar indicates the cleaning steps:
▶
Wait for the cleaning programme to be completed.
ð
The following message is displayed after completion of the cleaning programme:
▶
After completion of the "Weekly" cleaning programme, place hoses in holders.
6.7 Intensive germ reduction during prolonged periods of non-
use
During extended periods of non-use, e.g. semester breaks at university, "Weekly" clean-
ing must be run during the time of non-use. The germ reduction liquid remains inside the
system during the period of non-use.
Before the period of non-use
▶
Start the "Weekly" cleaning programme and switch it off during the active phase after
an acoustical signal within the dwell time of 30 minutes.
▶
To prevent damage due to hoses getting kinked, remove hoses from the hygiene
centre and place them in the dentist holder.
▶
Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶
Turn on the central compressor.