![KaVo ELECTROmatic TM Instructions For Use Manual Download Page 83](http://html1.mh-extra.com/html/kavo/electromatic-tm/electromatic-tm_instructions-for-use-manual_3703148083.webp)
Instructions for use ELECTROmatic TM and TMM/TMC
10 Troubleshooting
Malfunction
Cause
Remedy
Event E8
Saved data or settings
have been reset to
starting value.
▶
Confirm message and check or correct the pro-
gram settings.
▶
If error persists, notify the service engineer.
Event E12
No connection to mo-
tor.
▶
Check/restore correct connection.
▶
Replace motor cord.
▶
If error persists, notify the service engineer.
Event E14
Motor overload.
▶
Relieve the motor, stop and start on foot con-
trol.
All other events
Internal system error.
▶
Turn the unit off and on.
▶
If error persists, notify the service engineer.
83 / 90
Summary of Contents for ELECTROmatic TM
Page 1: ...Instructions for use ELECTROmatic TM and TMM TMC...