Instructions for use ELECTROmatic TM and TMM/TMC
6 Operation | 6.4 PREP mode
60 / 90
ð
SAFEdrive is active = Control element has a blue background.
ð
SAFEdrive is deactivated = Control element has a grey background.
ð
The settings are always saved instantaneously.
ð
In all activated speed memories, the SAFEdrive icon is shown with a blue
background.
Use with the SAFEdrive function
With regard to handpieces from third-party manufacturers, KaVo recommends
inactivating the SAFEdrive function as the SAFEdrive protection function is ef-
fective only with high-speed handpieces from KaVo.
See also:
2
6.4.3.1 Turning SAFEdrive on / off, Page 59
If SAFEdrive is activated and the handpiece is overloaded, the first stage,
"SAFEdrive alert", is activated:
▪ The motor power is reduced (can lead to a drop in motor speed).
▪ The LUX light starts to pulse.
▪ The SAFEdrive icon flashes yellow.
Requirement
The handpiece is not defective
▶
If a handpiece from a third-party manufacturer is used, deactivate SAFEd-
rive or use a high-speed handpiece from KaVo.
▶
If the SAFEdrive icon flashes yellow, relieve the handpiece of its load for at
least 2 seconds, i.e. lift it off the tooth.
ð
The motor reaches maximum speed again and the SAFEdrive icon turns yel-
low again.
▶
As soon as the motor reaches maximum speed again and the SAFEdrive
icon turns yellow, continue working as usual making sure to use intermit-
tent mode as the operating mode.
See also:
2
3.6 Technical specifications of the ELECTROmatic, Page 27
If the exposure to the load persists for more than 10 seconds, the stage,
"SAFEdrive error", is reached:
▪ The motor is stopped automatically.
▪ The SAFEdrive icon flashes red.
▶
If a handpiece from a third-party manufacturer is used, deactivate SAFEd-
rive or use a high-speed handpiece from KaVo.
▶
If an automatic motor stop signals that the "SAFEdrive error" status is evid-
ent, remove the handpiece from the mouth of the patient and proceed as
follows:
▶
Carefully check the handpiece head for:
- Temperature
- Damage
- Ability to rotate the dental bur
▶
The motor can be restarted if there is no damage or overheating evident.
To do so, press the foot control again.
▶
When you continue working make sure not to touch the cheek with the
handpiece head and, in particular, with the press-button.