
Instructions for use DSEclinical Typ 5198
5 Operation | 5.6 Using instruments
62 / 100
Adjustable, as an option:
▶
Take the instrument from the holder.
ð
This activates the cold light.
or
▶
Only if the instrument is connected to the dentist unit,
take the instrument from the holder and press the foot pedal.
ð
This activates the cold light.
Replacing the lamp
CAUTION
Danger of injury from a hot valve body.
Risk of burn injury.
▶
Switch main device switch off.
▶
Allow the instrument to cool down after extended use.
▶
Pull off the grip sleeve
④
together with the cannula from the valve body
①
.
Replacing the high-pressure lamp
▶
Push the holder
③
towards the front and draw and remove the defective
high pressure lamp
②
from its holder.
Summary of Contents for DSEclinical 5198
Page 1: ...Instructions for use DSEclinical Typ 5198...
Page 98: ......
Page 99: ......
Page 100: ...1 007 8663 Fk 20191001 3 en...