Instructions for use DSEclinical Typ 5198
5 Operation | 5.4 Using function through the dentist or assistant unit
WARNING
Wrong handling.
Reversible blinding (temporary sight impairment).
▶
Do not direct the light field at patients, users or/and third parties while you
switch on the light.
▶
Do not direct the light field at the patient's eyes when you move the light
head.
▶
Maintain a clearance of approx. 700 mm between the operating light and
the mouth of the patient.
CAUTION
Stroboscopic effect of the rotating instrument.
A stroboscopic effect could arise in instruments rotating at a certain speed dur-
ing application of the KaVoLUX 540 LED. This is an optical illusion that makes
the handpiece appear to be standing still or rotating extremely slowly.
Risk of injury.
▶
If the stroboscopic effect is evident, change the speed slightly and continue
working as usual.
CAUTION
Faulty measurement in connection with KaVo DIAGNOdent.
Simultaneous application of the operating light KaVoLUX 540 LED and the
KaVo DIAGNOdent can lead to faulty measurements.
▶
Switch the operating light to laser mode when using the
KaVo DIAGNOdent.
▶
Or switch off the operating light, do not use KaVo DIAGNOdent and operat-
ing light KaVoLUX 540 LED simultaneously.
CAUTION
Premature hardening of composite fillings.
A light intensity that is too high can have a negative impact on the durability of
the treatment.
▶
Select the appropriate dimming level according to the processing time.
Note
The KaVoLUX 540 LED T cannot be operated through the control functions of
the treatment centre. The operating light an only be switched on and off.
Turning the operating light On and Off
The KaVoLUX 540 LED operating light can be used in the following modes:
▪ Normal light: preset at 5,500 Kelvin and 30,000 Lux corresponding to day-
light quality
▪ COMPOsave mode: Enables longer working times for composites by filtering
the blue components of the light (compatible mode for light-activatable res-
toration materials).
▪ Dimmed light: approx. 4,000 Kelvin; equivalent to the light of a halogen op-
erating light
▪ Laser mode: Light mode with no detrimental effect on the KaVo KEY Laser
III, the KEY Laser 3+ or the KaVo DIAGNOdent
▶
Press the "Operating light" button on the dentist or assistant element.
49 / 100
Summary of Contents for DSEclinical 5198
Page 1: ...Instructions for use DSEclinical Typ 5198...
Page 98: ......
Page 99: ......
Page 100: ...1 007 8663 Fk 20191001 3 en...