23
ARCUS®digma Handheld.
A 8 Programming the articulator
with calculable values
A 8.1 Adjusting the Horizontal
Condylar Inclination HCI
Open the fixing devices for condylar box
right ” and left #, and rotate the condy-
lar boxes @ and £ to the desired setting.
Lock the fixing devices ” and #. The
values are legible on the edges fi.
A 8.2 Adjusting the Bennet Angle
Loosen the fixing devices for Bennet
Angle right \ and left Ì, then adjust the
condylar box inserts · and ¯ to the
desired setting.
Lock the fixing devices \ and Ì. The
settings are legible on the inner markings
|.
A 8.3 Adjusting the Immediate Side
Shift ISS
Loosen the fixing devices for Bennet
Angle right \ and left Ì, and set the
adjustment rings ‰, „ for Immediate Side
Shift to the required values ».
Lock the fixing devices \ and Ì.
Adjustment and reading of the values
ensues via the adjustment rings ‰, „ for
ISS.