Instructions for use Treatment light 9010 for MASTERspace
®
and ERGOspace
®
4 Assembly and start-up
|
4.4 Install and connect to equipment workplaces
Mount the light stand
▶ Screw the baseplate ⑩ and clamping plate ⑦ onto the worktop. Insert two
threaded bolts ⑥ through 11 mm dia. holes, and screw them tight underneath
the worktop.
▶ Fasten the connecting element ⑨ to the baseplate ⑩.
▶ Insert two connecting sleeves ③ in the light stand ⑧ and align them so that the
slots ① are vertical.
▶ Mount the light stand ⑧ on the connectors ⑨ and affix it with hexagon socket
screws ②.
▶ Vertically align the light stand ⑧ with adjusting screws ⑤.
▶ Mount the plastic covers ④.
Fasten the support arm with the light housing to the light stand
See also: 4.3.1 Fasten the support arm to the light housing with the height adjust‐
33/45