Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.9 Implant position
5.9 Implant position
The implant position can be entered as a two-digit number. The entry can be made
either in the FDI (ISO system) or in the Universal Numbering System (American Sys-
tem).
The implant position can be entered in any step of the program and is recorded if doc-
umentation is enabled.
See also:
2
5.2.2 Setting the documentation, Page 32
5.10 Documentation
The documentation function of the MASTERsurg LUX Wireless allows the date, time,
torque, speed, direction of rotation, program name, implant position, activity, transmis-
sion ratio, pump level, software version and serial number of the unit to be recorded.
Three modes are available:
▪ Implant: A torque plot is shown in the "Set implant" activity after each motor stop.
The plot visualises the torque profile applied during insertion of the implant (yellow
line). The green line describes the maximum torque reached. The torque plot is
also stored in the <IMPxxxxx.BMP> file on the SD card. All other data are recor-
ded as numerical values in the file, <IMPxxxxx.CSV>.
▪ Permanent: Documentation of the entire treatment process as numerical values in
the <REC000xx.CSV> file on SD card without BMP
▪ Off: Documentation disabled. No recording or output of the treatment data.