7 Processing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never disinfect the handpiece with chlorine-containing products.
Functional damage and property damage.
▶
Only disinfect in a washer disinfector or manually.
NOTICE
Never process this medical device in an ultrasonic device.
Functional damage and property damage.
▶
Clean manually or in a washer disinfector only.
7.3.1 Automated internal and external cleaning and
internal and external disinfection
KaVo recommends washer disinfectors in accordance with ISO 15883-1, which
are operated using alkaline cleaning agents.
▶
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the washer disinfector.
7.3.2 Automated drying
The drying procedure is usually part of the cleaning program of the washer dis-
infector.
Note
Please comply with the instructions for use of the washer disinfector.
▶
In order to prevent impairment of the KaVo medical device, make sure that
the inside and outside of the device is dry after the end of the cycle.
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
7.4 Care products and systems - Servicing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
CAUTION
Improper service and care.
Risk of injury.
▶
Perform regular proper care and servicing.