Instructions for use SMARTtorque S609 C - 1.008.1644 SMARTtorque Mini S605 C - 1.008.1645 SMARTtorque
S608 C - 3.001.6560
7 Processing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
7 Processing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection risk.
▶
Take suitable personal protective measures.
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
▶
Process the medical device as soon as possible after treatment.
▶
To minimize the risk of infection during processing, always wear protective
gloves.
▶
Remove the tool from the medical device.
▶
Remove all residual cement, composite or blood immediately.
▶
Do not immerse in solutions or the like.
7.2 Manual processing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
NOTICE
Never process this medical device in an ultrasonic device.
Functional damage and property damage.
▶
Clean manually or in a washer disinfector only.
7.2.1 Manual external cleaning
Accessories required:
▪ Tap water
▪ Brush, e.g. medium-hard toothbrush
▶
Brush under flowing tap water.
17 / 26