4.3
Checking the pressures
CAUTION
Compressed air connection on devices.
Dirty and moist compressed air causes premature bearing wear.
▶ Ensure that the cooling air is dry, clean and uncontaminated in accordance
with EN ISO 7494-2.
▶ Insert the test gauge (Mat. no. 0.411.8731) between the
MULTIflex (LUX) / MULTIflex LED coupling and the medical device, and check
the following pressures:
- Drive pressure: 2.8 bar (40 psi)
- Return air: < 0.5 bar (7 psi)
- Water: 0.8 to 2.5 bar (12 to 36 psi)
4.4
Checking the O-rings
CAUTION
Missing or damaged O-rings.
Malfunctions and premature failure.
▶ Make sure that all O-rings are on the coupling and undamaged.
Number of available O-rings: 5
Instructions for use SMARTtorque LUX S619 L - 1.008.1641 SMARTtorque Mini LUX S615 L - 1.008.1643
SMARTtorque S619 C - 1.008.1642
4 Start up and shut down | 4.3 Checking the pressures
8/21