background image

Instructions for use

EXPERTmatic LUX E15 L – 1.007.5530 
EXPERTmatic LUX E20 L – 1.007.5540 
EXPERTmatic E15 C – 1.007.5531 
EXPERTmatic E20 C – 1.007.5541

Summary of Contents for 1.007.5530

Page 1: ...Instructions for use EXPERTmatic LUX E15 L 1 007 5530 EXPERTmatic LUX E20 L 1 007 5540 EXPERTmatic E15 C 1 007 5531 EXPERTmatic E20 C 1 007 5541...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...3 Inserting the dental bur 16 5 4 Removing the dental bur 17 6 Checking for malfunctions and troubleshooting 18 6 1 Checking for malfunctions 18 6 2 Troubleshooting 18 6 2 1 Replacing the O rings on t...

Page 4: ...aVo Original Factory Repair using www kavobox com KaVo Technical Service If you have any questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo co...

Page 5: ...and storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original l...

Page 6: ...e EXPERTmatic LUX E15 L 1 007 5530 EXPERTmatic LUX E20 L 1 007 5540 EXPERTmatic E15 C 1 007 5531 EXPERTmatic E20 C 1 007 5541 1 User instructions 6 32 CAUTION In cases which if not prevented can lead...

Page 7: ...ure is effective 4 Reprocess the product and accessories appropriately before disposal 4 In the case of injury to soft tissue do not continue treatment in the oral cavity with instruments driven by co...

Page 8: ...e motors regularly 4 Do no use any damaged motors 4 Do not use motors for unauthorised purposes 2 4 Accessories and combination with other equipment Use of non authorised accessories or non authorised...

Page 9: ...year 4 KaVo recommends specifying in house service intervals where the medical device is brought to a professional shop for cleaning servicing and a func tion check Define the service interval depend...

Page 10: ...c E20 C 1 007 5541 3 Description of the product 10 32 3 Description of the product EXPERTmatic LUX contra angle handpiece E15 L Mat no 1 007 5530 EXPERTmatic LUX contra angle handpiece E20 L Mat no 1...

Page 11: ...with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions for use According to these regulations the user is required to only u...

Page 12: ...to 9 5 Nl min Pushbutton chuck With Pushbutton chuck Insert Dental burs and diamond grinders in accordance with DIN EN ISO 1797 type 1 Can be attached to All INTRA LUX motors and motors with a connect...

Page 13: ...tely in accordance with current packaging law employing waste management recycling companies Comply with the comprehensive return system KaVo has had its packaging licensed for this purpose Please com...

Page 14: ...ructions for use EXPERTmatic LUX E15 L 1 007 5530 EXPERTmatic LUX E20 L 1 007 5540 EXPERTmatic E15 C 1 007 5531 EXPERTmatic E20 C 1 007 5541 4 Startup and shut down 4 1 Checking the amount of water 14...

Page 15: ...at least 2 minutes without transmission handpieces being at tached and if there is a risk of contamination from reflux or back suction the system may also need to be rinsed for 20 to 30 seconds after...

Page 16: ...SO 1797 type 1 are made of steel or hard metal and meet the following criteria Shaft diameter 2 334 to 2 350 mm Overall length max 22 mm WARNING Use of unapproved dental burs Injury to the patient or...

Page 17: ...nd chuck system reduction of the service life of dental bur and chuck system 4 Do not operate the dental bur at a higher speed than recommended by the manufacturer 4 Firmly press the push button with...

Page 18: ...the O ring Also refer to Instructions for use of motor 6 2 Troubleshooting WARNING Repair WITHOUT using KaVo original spare parts Parts such as the cover can come loose and cause injury Aspiration swa...

Page 19: ...gers to form a loop 4 Push the O ring to the front and remove it 4 Insert new O rings into the recesses and spray them with KaVo Spray 6 2 2 Cleaning the spray nozzle CAUTION Overheating of the tooth...

Page 20: ...ing always wear protect ive gloves 4 Remove the dental bur from the medical device 4 Remove all residual cement composite or blood immediately 4 Do not place in solutions or similar substances 7 2 Man...

Page 21: ...ss the medical device with chloride containing products Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only KaVo recommends the following products based on the compatib...

Page 22: ...ns for use of the disinfectant 7 2 5 Manual drying Use KaVo DRYspray for subsequent drying of the air water and gear unit ducts 4 Cover the medical device with the KaVo Cleanpac bag and place it on th...

Page 23: ...for Use of the washer disinfector 4 For adaptations unless otherwise specified refer to the Instructions for Use of the washer disinfector 7 3 2 Automated drying The drying procedure is usually part...

Page 24: ...are no validated cleaning of the interior in accordance with German Robert Koch Institute requirements KaVo recommends servicing the product as part of the reprocessing after each use i e after each c...

Page 25: ...a sterile pack 7 6 Sterilisation Sterilisation in a steam steriliser autoclave in accordance with EN 13060 EN ISO 17665 1 CAUTION Improper service and care Risk of injury 4 Service regularly with suit...

Page 26: ...X E15 L 1 007 5530 EXPERTmatic LUX E20 L 1 007 5540 EXPERTmatic E15 C 1 007 5531 EXPERTmatic E20 C 1 007 5541 7 Reprocessing steps in accordance with ISO 17664 7 7 Storage 26 32 NOTE Comply with the e...

Page 27: ...ised dental dealers Material summary Mat no INTRA Instrument stand 3 005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle needle 0 410 0921 KaVo CLEANspray 2110 P 1 007 0579...

Page 28: ...andatory statutory provisions KaVo shall not be liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non com...

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ...1 008 2446 bd 20221208 03 en...

Reviews: