Instructions for use EXPERTmatic LUX E15 L – 1.007.5530, EXPERTmatic LUX E20 L – 1.007.5540, EXPERTmatic
E15 C – 1.007.5531, EXPERTmatic E20 C – 1.007.5541
5 Operation | 5.2 Removing the medical device
16 / 32
4
Lightly spray O-rings on the motor coupling with KaVo Spray.
4
Place the medical device on the motor coupling and lock it into place.
4
Pull on the medical device to make sure that it is securely affixed to the
coupling.
5.2 Removing the medical device
NOTE
Do not pull the medical device off the motor coupling while holding it by the
instrument head.
4
Unlock the medical device from the motor coupling by twisting it slightly and
then pulling it along its axis.
5.3 Inserting the dental bur
NOTE
Only use carbide burs or diamond burs that conform to DIN EN ISO 1797 type
1, are made of steel or hard metal, and meet the following criteria:
- Shaft diameter: 2.334 to 2.350 mm
- Overall length: max. 22 mm
WARNING
Use of unapproved dental burs.
Injury to the patient or damage to the medical device.
4
Comply with the instructions for use and the intended use of the dental bur.
4
Only use dental burs that do not deviate from the specified data.
CAUTION
Do not use damaged dental burs.
Risk of injury from swallowing a dental bur that falls out.
4
Do not use damaged dental burs.
4
Do not use dental burs that have been hit.
4
Do not use dental burs with adherent soiling.
4
Do not use visibly imbalanced dental burs.
CAUTION
Contaminated, sharp-edged dental bur.
Infections or cuts.
4
Use gloves or a finger guard when you test, insert and remove the dental
bur.
CAUTION
Dental bur with damaged, worn or deformed shafts.
Injury hazard, dental bur may fall out during treatment.
4
Never use dental burs with damaged, worn or deformed shafts.