Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390
7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing
KaVo recommends the following products based on material compatibility. The
microbiological efficacy must be ensured by the disinfectant manufacturer.
Approved disinfectants:
▪ CaviWipes and CaviCide made by Metrex
▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)
▪ FD 322 made by Dürr
Required tools:
▪ Cloths for wiping off the product.
▶
Spray the disinfectant on a cloth then wipe the product and let it work ac-
cording to the disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
7.2.4 Manual disinfection - internal
Not applicable.
7.2.5 Manual drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
7.3 Automated reprocessing
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
For manual cleaning only.
NOTICE
Malfunctioning from using a disinfectant bath or chloride-containing
disinfectants.
Defects in the device.
▶
Disinfect manually only!
7.3.1 Automated internal and external cleaning and
internal and external disinfection
Not applicable.
7.3.2 Automated drying
Not applicable.