KATflow 200
5 Commissioning
Output
Display
Top Line
Units (Select from list
↑↓
)
Middle Line
Units (Select from list
↑↓
)
Bottom Line
Units (Select from list
↑↓
)
Damping
Reduces fluctuations in the display output.
1 ... 255 s
Datalogger
Interval
0 ... 999 s
Selection
From list. <ENTER> selects, <0> deselects.
Up to ten variables may be logged.
Low memory
Warning output
0 ... 100 %
Log Wrap
Saves "selected" items as a continuous stream
without headers (Note : this means files cannot
be processed by )
Yes/No
Log download
Sends all log data using RS232 serial port
Log erase
Clears the logger
Serial
communica-
tion
Mode
Select from list
↑↓
None
Printer
(output every second of selected values)
Diagnostic
Download
(send logger data using RS232)
Cal Test
(laboratory calibration, not recommen-
ded for field or customer use)
Baud
Select from list
↑↓
9600
(Default)
19200,
57600,
115200
Parity
Select from list
↑↓
None
Even
(Default)
Odd
System
Instrument info
Model code
KF200
Serial number
Example:
20000003
HW revision
Example:
2.0, 1.5
SW revision
Example:
3.2, 3.1
Calculation
Low flow cut off 0 ... 0.10 m/s
Max. flow cut
off
0 ... 30 m/s
Corrected
Yes (flow profile correction)
No
PV offset
-30 ... 30 units
PV scaling
0 ... 1000 units
Zero calibration Adjust :
Zero (Yes/No) : Sets current flow as zero
Track (Yes/No) : Zero follows output variations
Delta time : Zero flow offset in ns
Time up : Upstream offset in ns
28