OPERATING AND HANDLING
BRAVA STRETCHER
7
Rev. 9
20.12.2022
- The use of accessories, transducers, and cables other than those specified, except for transducers and cables sold by
Kartsana as replacement parts for internal components may result in increased emissions or decreased immunity of the TG-
1000 BRAVA stretcher.
CAUTIONS:
- Changes or modifications to the unit not expressly approved by Kartsana
could void the user’s authori
ty to operate the system.
- The stretcher can be set at any stretcher load height position
.
Establish the required stretcher load height before placing the
stretcher into service
- Set the stretcher load height to the proper stop height prior to operation.
- Installation of the safety hook should be done by a certified mechanic familiar with ambulance vehicle construction. Consult the
vehicle manufacturer before installing the safety hook and be sure that the installation of the safety hook does not damage or
interfere with the brake lines, oxygen lines, fuel lines, fuel tank or electrical wiring of the vehicle.
- Only use the battery and charger as specified in this Instruction Manual.
- The stretcher is not for use with an AC adapter.
- When charging a battery in an ambulance, locate the charger in an enclosed cabinet and out of patient reach during transport.
- Ensure that the battery is fully charged prior to placing into service
.
An uncharged or depleted battery may cause poor
stretcher performance.
- Before operating the stretcher, clear any obstacles that may interfere and cause injury to the operator or patient.
- When unloading the stretcher from the patient compartment, ensure that the caster wheels are safely set on the ground or
damage to the product may occur.
- Remove the battery if the stretcher is not going to be used for an extended period (more than 24 hours).
- Wheel lock(s) are only intended to help prevent the stretcher from rolling while unattended and to aid in patient transfer. A
wheel lock may not provide sufficient resistance on all surfaces or under loads.
- Ensure that the restraints are not entangled in the base frame when raising and lowering the stretcher.
- Do not store items under the stretcher mattress
.
Storing items under the mattress can interfere with the operation of the
stretcher.
- Do not steam clean or ultrasonically clean the unit.
- Maximum water temperature should not exceed 80°C.
- Allow stretcher to air dry.
-
Towel dry all casters and interface points.
- Failure to comply with these instructions may invalidate any/all warranties.
- Always remove the battery before washing the stretcher.
- For additional maintenance information, see the preventative maintenance information.
- Improper maintenance can cause injury or damage to the product
.
Maintain the stretcher as described in this manual
.
Use only
Kartsana approved parts and maintenance procedures
.
Using unapproved parts and procedures could cause unpredictable
operation and/or injury and will void the product warranty
- Failure to use authorized parts, lubricants, etc
.
could cause damage to the stretcher and will void the warranty of the product.
- Hydraulic lines, hoses, and connections can fail or loosen due to physical damage, kinks, age, and environment exposure
.
Check hoses and lines regularly to avoid damage to the stretcher
.
Check and tighten loose connections.
NOTES:
- Loose items or debris on the patient compartment floor can interfere with the operation of the safety hook and stretcher
fastener
.
Keep the patient compartment floor clear
- This manual should be considered a permanent part of the stretcher and should remain with the product even if the stretcher is
subsequently sold.
Summary of Contents for BRAVA TG-1000
Page 2: ...OPERATING AND HANDLING BRAVA STRETCHER Rev 9 20 12 2022...
Page 12: ...OPERATING AND HANDLING BRAVA STRETCHER 10 Rev 9 20 12 2022 FOLDED LEGS...
Page 38: ...OPERATING AND HANDLING BRAVA STRETCHER 36 Rev 9 20 12 2022 5 EXPLODED STRETCHER BRAVA TG 1000...
Page 39: ...OPERATING AND HANDLING BRAVA STRETCHER 37 Rev 9 20 12 2022...
Page 40: ...OPERATING AND HANDLING BRAVA STRETCHER 38 Rev 9 20 12 2022 5 1 Complete paciente area BRAVA...
Page 41: ...OPERATING AND HANDLING BRAVA STRETCHER 39 Rev 9 20 12 2022...
Page 42: ...OPERATING AND HANDLING BRAVA STRETCHER 40 Rev 9 20 12 2022 5 2 Chassis BRAVA...
Page 43: ...OPERATING AND HANDLING BRAVA STRETCHER 41 Rev 9 20 12 2022...
Page 44: ...OPERATING AND HANDLING BRAVA STRETCHER 42 Rev 9 20 12 2022 5 3 Doble leg brace BRAVA...
Page 46: ...OPERATING AND HANDLING BRAVA STRETCHER 44 Rev 9 20 12 2022 6 ACCESORIES BRAVA TG 1000...
Page 47: ...OPERATING AND HANDLING BRAVA STRETCHER 45 Rev 9 20 12 2022...