12.06 / V1.0
3-7
3.4 Troubleshooting
3.4.1 Error codes
AUTOCON
®
II 400
MODEL 205352 20-1
Error code
Description of error
Solution
Priority
A-02
Communication error during activation
High
A-0A
NE module not available
High
A-0B
NE contact: resistance too low
Check neutral electrode and reposition
if necessary
Medium
A-0C
NE contact: resistance too high
Check neutral electrode and reposition
if necessary
Medium
A-0E
Erroneous ESI communication
High
A-0F
24 V supply voltage too low
High
A-10
Release activation (Termination by
autostop)
End activation of HF energy
Low
A-11
Missing activation signal
Send instrument to service technician
High
A-12
Activation during power up not allowed
Check footswitch, handpiece and cable
Medium
A-17
Simultaneous activation not permitted
Activate either via footswitch or
handpiece
Medium
A-18
Missing deactivation signal from HF
module
High
A-19
Power fail signal detected
Check power cord, power outlet and
power supply
Low
A-1C
Max. activation time exceeded
Cancel activation
Low
A-1E
Activated touch screen during power up
Touch screen must not be pressed
during power-up
Low
A-1F
Critical NE symmetry (activation
terminated)
Check neutral electrode and reposition
if necessary
High
A-22
Remove tissue contact, Autostop
reached
Break contact with tissue
Low
A-23
Erroneous message from HF module
High
A-25
Illegal socket module configuration
Low
A-50
Incompatible ESI version
Medium
A-51
Incompatible Black Box version
Medium
A-52
Internal device temperature out of
specs (> 70 ˚C )
High
A-53
Low capacitance of back-up battery
Low
A-55
Back-up battery empty (no data
storage)
Low
A-56
Procedure level B not existent
Low
A-60
Procedure transfer not possible (SCB)
Low
A-81
Illegal ESI message
High
A-84
Dual-pedal footswitch recognized
Low
A-85
Dual-pedal footswitch disconnected
Reconnect dual-pedal footswitch and
check cable
Low
A-88
Single-pedal footswitch recognized
Low
A-89
Single-pedal footswitch disconnected
Reconnect single-pedal footswitch and
check cable
Low
A-8E
Missing status message from HF socket
1
High
A-8F
Missing status message from HF socket
2
High
A-90
Missing status message from HF socket
3
High
A-93
Triple-pedal footswitch recognized
Low
A-94
Triple-pedal footswitch disconnected
Reconnect triple-pedal footswitch and
check cable
Low
Summary of Contents for AUTOCON II 400
Page 8: ......
Page 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x...
Page 10: ......
Page 13: ...III Ger teabbildungen Photographs of the equipment Im genes del equipo...
Page 14: ...IV W R Q Ger teabbildungen Photographs of the equipment Im genes del equipo I O Y U T E...
Page 146: ......
Page 152: ......
Page 156: ......
Page 158: ......
Page 172: ......
Page 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1...
Page 192: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 203: ......