"Loc" is displayed in the hot air temperature section, and
"Lc1" is displayed in the set temperature section.
Remove the cap of the external sensor socket
and plug in the external sensor KX350AP (the
external sensor is an automatic setting type by
mounting).
When the temperature of the constant temperature bath is controlled using an external sensor, the multi-
dryer is stopped when the temperature inside the constant temperature bath reaches the set temperature,
and the temperature inside the tank is -5 ° C to -10 ° C (optional). Once the setting is reached, you can
perform energy-saving operation to restart the operation of the multi-dryer.
Set the hot air temperature while stopped.
Since this set temperature is the upper limit temperature
of the discharge port of the multi-dryer, set the
temperature in consideration of the set temperature of
the external heat dissipation from the piping,
etc. (If there is no upper limit on the discharge port
temperature, set it to 300 ° C). ..
The external sensor lamp (orange) lights up and "0" is
displayed in the set temperature section.
※Please make changes with the external
operation function turned off.
4.Intermittent operation
Set the temperature of the place controlled by
the external sensor.
Press and hold the hot air switch and hot air
temperature up key for about 3 seconds.