26
Chapter 7 Main Specification
MODEL
6750
Measuring
Range
Airflow
±8 to 600 m³/h
Temperature 0 to 50
℃
Measuring
Accuracy
Airflow
8
~
350 m
3
/h
︓
±
(
Reading x 3%
+
1digit m
3
/h
)
350 to 600 m
3
/h
︓
±
(
Reading x 5%
+
1digit m
3
/h
)
Temperature ±0.5
℃
Resolution
Airflow
1 m
3
/h
Temperature 0.1
℃
Operating Environment
0 to 50
℃
(
No condensation
)
Storage Environment
-10 to 50
℃
(
No condensation
)
Atmospheric pressure
compensation
By a manual setting of atmospheric pressure value
Data Storage
6,000 data
Communication
Bluetooth
®
、
Digital Output: USB
Power Supply
AA Battery x 4, AC adaptor (Option)
Dimension of the hood
355×355 mm
Weight
1.5 kg
Accessories
Indicator, Indicator Fixing Jig, Extension Rod,
Carrying Case, Hood (355×355 mm),
USB Cable, Main Body-Indicator Cable (2m),
AA Batteries (4 pieces), Measuring software (CD-
R), User's Manual (this document),]Insp
document), Inspection Report
Options
Spare Hood (6750-01), Hood Edge Urethane
Rubber (6750-02), Hood support pole (6750-07),
Main Body-Indicator Cable (6750-08), USB
Communication Cable (6750-09), Measuring
Software (6750-40), Printer (NP-DPU-S245),
Printer Roll Paper (TP-202L), Printer Cable(6750-
10), Carrying Case (6750-B1), AC adaptor (Model
3887-01), Traceability