9 Technical service and maintenance 

 

 

NOTE: All spare parts used for replacement or repairs shall be supplied by KANMED or its authorised 
distributor. Use of “non KANMED permitted parts” automatically transfers all safety responsibilities to 

the hospital. Only qualified personnel shall attempt repairs. 

 

9.1 Annual checks 

 

Check the following at least once a year: 

 

That cables and connectors free of defects. 

 

Test the foldable sides for cracks or damages. 

 

That the switch for pillar height control responds without delay, that the raising and lowering 
function is working correctly and that it makes no noise when operated. 

 

That the tilt function works correctly. It should not make any noises. 

 

That the wheels are tightened to the stand and not damaged in any way, that they roll without 

noise, and that the wheels locking function works properly. 

 

That all screws are properly tightened, especially the screws connecting the height adjusting 

pillar to the wheel base, the screws on top of the pillar, the screws holding the shelf and gas 

spring, the clips holding the bolt for the bed surface.  

 

9.2 Adjusting the release of the gas spring 

 

By manoeuvring the release arm you press down the release pin at the top of the gas spring. Adjust by 

releasing the gas spring from the pillar. Start by taking away the small safety spring. Make someone or 

something hold the bed. Pull out the gas spring from the pillar. Un-tighten the locking nut by turning 
the gas spring you adjust the contact with the release arm.  

 

9.3 Changing side walls 

 

The easiest way to replace a side wall is to place it in its hanging down position, insert one pin in the 
lower grove, bend the wall and insert the other pin. 

 

9.4 Changing the side brackets

 

 

The side wall brackets are screwed from the inside of the bed with two (2) screws at each corner. 

 

9.5 Changing the electrical control 

 

Always unplug the power before doing repair on the Babybed. 

 

Remove the two (2) Torx screws under the power inlet and slide the power supply out of the 

box and detach the spiral cable from the connectors. 

 

Remove the two screws (Philips) under the remote control and slide the unit by pulling it gently 
out from the front of the bed. 

 

9.6 Changing the pillar 

 

Disconnect power. Remove the two (2) Torx screws under the power inlet and slide the power supply 

out of the box and detach the spiral cable and the pillar cable from the connectors. Loosen the gas 
spring as described in 9.2. Remove the pin that holds the bed to the pillar and take away the bed. 

Remove the bed holder from the top of the pillar. Loosen the 6 screws that hold the pillar to the wheel 

base. Mount the new pillar. Connect the cables and test run the new pillar. Rebuild the bed again. Make 
sure that all screws are properly tightened.  

Summary of Contents for Babybed

Page 1: ...User Manual Art no BB1 070 1 March 2007 Manufactured by KANMED AB www kanmed se Distributed by Caution Please read this manual carefully Incorrect use may cause patient injury...

Page 2: ...s 3 3 Preparations before use 4 4 Handling and daily use 5 5 Cleaning and Maintenance 6 6 Trouble shooting 6 7 Warranty Statement 6 8 Technical Data 7 9 Technical service and maintenance 8 10 Spare pa...

Page 3: ...weight limits Do not place electronic equipment on the shelf under the bed 2 Description of Symbols Maximum Load permitted Bed surface for BB 1 and BB 3 10 kg Excl BABYWARMER Control Unit 3kg and Wate...

Page 4: ...nt has a channel inside Thread the Tent pole into the channel in order to secure it Place the mattress in the bed 3 4 Check all bolts and screws Make sure that all bolts and screws are properly tighte...

Page 5: ...for BB 3 The height is adjustable by the pushbuttons on the front The lifting column has an overload protection that will not be activated under normal conditions Constantly adjusting height with heav...

Page 6: ...not keep the position Something is blocking it Check that nothing blocks it Bed feels unstable Loose screws or wheels Technical service Height adjustments makes unusual noises Technical service 7 War...

Page 7: ...I Class I NA Used standards SS EN 60 601 SS EN 60 601 Products carries the CE mark Conforms to UL 60601 1 Certified to CAN CSA C22 1 No 601 1 EMC The Kanmed Babybed contains no electronics sensitive...

Page 8: ...ring from the pillar Start by taking away the small safety spring Make someone or something hold the bed Pull out the gas spring from the pillar Un tighten the locking nut by turning the gas spring yo...

Page 9: ...3 Complies The Babybed is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings u...

Page 10: ...tage prior to application of the test level Guidance and manufacturer s declaration electromagnetic immunity The Babybed is intended for use in the electromagnetic environment specified below The cust...

Page 11: ...d The Babybed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Babybed can help prevent electromagnetic interferen...

Page 12: ...KANMED AB G rdsfogdev gen 18B SE 168 66 BROMMA Stockholm Sweden Telephone 46 8 56 48 06 30 Telefax 46 8 56 48 06 39 E Mail info kanmed se Home page www kanmed se Distributed by...

Reviews: