User Manual BB100-400 /070
Page 6 of 20
4.5 Connecting the BabyBed to the mains
(not for BB300/301 and BB400F/401F)
The BabyBed must be connected to an
earthed power outlet using a grounded cable.
Route the cable so that no one can get caught
or fall because of it.
NOTE: The only way to break the power
is to either disconnect the cable at the
Pillar base or at the wall socket.
Place the Bed so that the cable
can easily be disconnected.
WARNING: To avoid the risk of
electrical shock, this equipment must
only be connected to a power outlet
that has protective earth.
4.6 Adjusting height
(not for BB300/301 and BB400F/401F)
The height is adjustable by pressing the Foot Pedal
at the wheel base.
Connect power and run the Bed up and down
and ensure that it moves smoothly without
any unusual noise.
The motor can be heard which is an
indication/warning that the Bed Platform is now moving.
Foot pedals can be mounted on the left or right side or on
both sides. Front pedal is also available.
WARNING:
When changing height, make sure that no person, cable or other items prevents the
movement and that no cable is jammed or wrapped between the Pillar and the Wheelbase or
somewhere else.
If children, relatives or patients are at risk of playing with the foot pedal, then disconnect the
power. Make sure that the Kanmed BabyWarmer still has power.
Do not block the movements when activating the electrical height adjustment.