Kane 505 Instruction Manual Download Page 3

3

FUNCTIONS

FEATURES

  1000V AC/DC

  10A AC/DC

  Resistance 40MΩ

  Diode test

  Audible continuity

  Capacitance 9999µF

  Milliamps/Microamps

  Auto/Manual ranging

  Auto power off

  Min/Max

  Hold

  1 ms Fast response Min/Max

  Low battery indicator

  Test lead holders

  Kick stand

  High resolution backlit display

  Auto selection

  Fused test lead inputs

  Rubber boot

GENERAL SPECIFICATIONS

• 

  Operating Temperature:  

32˚ to 122˚F (0˚ to 50˚C)

• 

  Storage Temperature:  

-44˚ to 122˚F (-20˚ to 50˚C)

• 

  Operating Humidity:  

<75% max.

• 

  Operating Altitude:  

6561 ft (2000m)

• 

  Display:    

6,000 count

• 

  Backlight:  

Yes

• 

  Over-range: 

“OL” is displayed

• 

  Dimensions: 

6.50” x 3.31” x  2.28” 

• 

  Item Weight:  

15.2 oz

• 

  Calibration:  

Recommended annually

• 

  CAT Rating: 

CAT III 600V/CAT II 1000V

• 

  Certifications:  

cETLus UL 61010-1:2012 3rd

CE EN 61010-1:2010 3rd, IEC61010-2-033:2012 Ed.1, EN 61326-1:2013

FCC, RoHS Compliant, TOV protection, IP 42. 6’ Drop protection

• 

  Battery Type:  

(AA)  2

• 

  Test Leads:  

CAT III test leads

• 

  Accuracy: 

± (% of r # of least significant digits)

Measurement Category

Short-Circuit (typical) kA

a

Location in the building installation

II

< 10

Circuits connected to mains socket outlets and similar 

points in the MAINS installation

III

< 50

Mains distributions parts of the building

IV

> 50

Source of the mains installation in the building

CATEGORY DEFINITIONS

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for 505

Page 1: ...FAST V Apo A A mA A A mA 10A MAX FUSED 400mA MAX FUSED MAX CAT III 600V CAT II 1000V INSTRUCTION MANUAL ENGLISH 1000V CAT II 600V CAT III KANE505 1000V Digital Multimeter KANE505 www com information...

Page 2: ...BOLS 6 OPERATING INSTRUCTIONS AC DC Voltage 1000V AC DC 6 Resistance 40M 7 Continuity 7 Diode 8 Capacitance 8 AC DC Microamps 6000 A 9 AC DC Milliamps 400mA 9 AC DC Amps 10A 10 Test Lead Notes 10 Batt...

Page 3: ...ge OL is displayed Dimensions 6 50 x 3 31 x 2 28 Item Weight 15 2 oz Calibration Recommended annually CAT Rating CAT III 600V CAT II 1000V Certifications cETLus UL 61010 1 2012 3rd CE EN 61010 1 2010...

Page 4: ...ys adhere to national and local safety codes Use proper personal protective equipment PPE to prevent shock and arc blast injury where hazardous live conductors are exposed Always turn off power to a c...

Page 5: ...ress and hold to return to live readings E Select Button Press to select AC DC or Auto Selection in the following functions Voltage A mA Amps F Function Dial Turns on meter and is used to select the f...

Page 6: ...s DC RMS at either the common or multifunction input ports as measured from earth ground Features RANGE MIN MAX HOLD SEL RANGE SEL MIN MAX HOLD FAST V Apo A A mA A A mA 10A MAX FUSED 400mA MAX FUSED M...

Page 7: ...tion Approx 1 0V Approx 40 600V Features MIN MAX HOLD Buzzer sounds at less than 40 WARNING Do not measure resistance on a live circuit Continuity RANGE SEL MIN MAX HOLD FAST V Apo A A mA A A mA 10A M...

Page 8: ...voltage drop Features RANGE MIN MAX HOLD AC VOLTS Ranges Accuracy Resolution Overload Protection 10nF to 9999 F 3 0 5 dgts 0 01nF to 1 F 600V Capacitance RANGE SEL MIN MAX HOLD FAST V Apo A A mA A A...

Page 9: ...2 5 dgts 0 01mA to 0 1mA 600mA 1000V Fast Fuse Minimum sensitivity 5mA AC auto selection mode only DCmA Ranges Accuracy Resolution Overload Protection 60mA to 400mA 1 0 3 dgts 0 01mA to 0 1mA 600mA 10...

Page 10: ...mA A Apo AT A Test Lead Notes CATIII 1000V Measurement Locations Ensure the test lead shield is pressed firmly in place Failure to use the CATIV shield increases arc flash risk CATII 1000V Measurement...

Page 11: ...communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can...

Page 12: ...age expenses or economic loss A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered Instruments out of warranty will be repaired when re...