KANE501C Manual
Page 5
ambient levels rotate the selector to “OFF” and switch off the analyser.
The instrument will count down from 10 to switch off.
CAUTION!
To optimise the measuring accuracy of the instrument, a zero should be
performed after the first 20 minutes of use (either with the pump running
or in the standby position). To do this, remove the probe from the
flue/exhaust port, or disconnect the hose from the gas inlet on the
analyser and turn the analyser off and on again in fresh air.
1.6 Post
Test
Remove the probe from the flue and allow the analyser to purge
with fresh air until readings return to zero. (
Be careful as the probe
tip will be HOT
)
Drain water trap.
Check particle filter.
WARNING!
This meter does not measure Carbon Monoxide (CO). Only measuring
CO
2
does not indicate whether a combustion process is running fuel rich
or fuel lean. A fuel rich combustion process can produce potentially
dangerous levels of CO so other measurements must be made to ensure
that the combustion process is running fuel lean and that the CO levels
are acceptably low.