4. Glossary 

 
4.1  Selector and Display Parameters on KANE501C 
 

2Er0

” – Analyser is performing the initial zero setting. 

 
CO

2

 (Carbon Dioxide) – Direct reading of the Carbon Dioxide Sensor 

displayed in percentage (%)  
 

Eff 

(Efficiency) – Calculated combustion (not appliance) efficiency based 

on measurements of CO

2

, Flue Temperature and fuel selected. The 

KANE501C calculates nett combustion efficiency and automatically 
determines if the combustion is in the condensing mode to make 
appropriate corrections. 
 

X-Air 

(Excess Air) – Excess Air is calculated for each fuel type from the 

CO

2

 measurement. It assumes that the combustion process is running 

fuel lean. It is important to make other measurements to confirm that this 
assumption is always correct. 

 
O

2

 (Oxygen) – O

2

 is calculated for each fuel type from the CO

2

 

measurement. It assumes that the combustion process is running fuel 
lean. It is important to make other measurements to confirm that this 
assumption is always correct. 
 

TFlue

 (Flue Temperature) – TFlue is a direct measurement of the 

temperature at the tip of the flue probe. This measurement is used to 
determine the net temperature for use in calculation of the combustion 
efficiency. 
 

TNet

 (Net Temperature) – Differential temperature of TFlue and ambient 

(or inlet if set) 
 

BAT

 symbol – Displays the Battery Power available. 

 
 
 
 
 
 

Summary of Contents for 501C

Page 1: ...KANE501C Performance Tester Stock No 18708 2 June 2012 Registered Community Design No 001051940 Kane International Ltd...

Page 2: ...ING 5 1 6 POST TEST 5 1 7 OVERVIEW 6 2 MAINTENANCE 7 9 2 1 GENERAL MAINTENANCE 7 2 2 PERIODIC SERVICE 7 2 3 ANNUAL RE CALIBRATION 7 2 4 CLEANING 7 2 5 WATER TRAP 8 2 6 CHANGING PARTICLE FILTER 8 2 7 B...

Page 3: ...due considerations of all the potential hazards 1 2 Pre test Checklist Clean particle filter Water trap and probe are empty of water All hose and thermocouple connections are properly secured Flue gas...

Page 4: ...d then it will complete the zero process NOTE If you wish to use the same fuel as previously selected rotate the selector to Stand By at power on The fuel in use will be displayed and then blink If th...

Page 5: ...nalyser and turn the analyser off and on again in fresh air 1 6 Post Test Remove the probe from the flue and allow the analyser to purge with fresh air until readings return to zero Be careful as the...

Page 6: ...KANE501C Manual Page 6 1 7 Overview Fuel Test Type Indicator Water Trap LCD Display Selector Dial...

Page 7: ...could result in physical degradation of the instrument This could alter the protection from personal injury this analyser provides to the operator Perform only those maintenance tasks that you are qu...

Page 8: ...ng the Particle Filter This is a very important part of the analyser and should be changed regularly It prevents dust and dirt particles from entering the pump and sensors that will cause damage The f...

Page 9: ...be appropriate to check the analyser prior to use The following procedure should be adopted Go through the normal start up sequence in the location where the equipment is to be used Switch on all loca...

Page 10: ...ctors Meter does not respond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Net temperature or Efficiency calculation incorrect Ambient tempe...

Page 11: ...fuel type from the CO2 measurement It assumes that the combustion process is running fuel lean It is important to make other measurements to confirm that this assumption is always correct O2 Oxygen O...

Page 12: ...KANE501C Manual Page 12 Fuels Available Selector Position Display Type L Oil LOIL Light Oil Propane PrOP Propane Nat Gas nGAS Natural Gas H Oil HOIL Heavy Oil Wood wOOd Wood Coal COAL Coal...

Page 13: ...y 2 0 1 0 110 Excess Air 3 0 1 0 250 Pre Programmed Fuels Natural Gas Light Oil Propane Wood Heavy Oil Coal Dimensions Approximate Weight Handset Probe 0 5 kg 175 x 45 x 80mm 6mm diameter x 250mm long...

Page 14: ...KANE501C Manual Page 14...

Page 15: ...KANE501C Manual Page 15...

Page 16: ...KANE501C Manual Page 16 Thank you for buying this instrument Before use please register on our website www kane co uk...

Reviews: