17
KANE 458s MANUAL
• Do not put the analyser on a hot surface
• Do not exceed the water trap’s levels
• Do not let the analyser’s particle filter become dirty and blocked
Look at your analyser’s displayed data to ensure stable operating
conditions are achieved & readings are within expected range.
CO PROTECTION PUMP OPERATION
Your analyser’s CO sensor is automatically protected from high levels of
CO. When levels of CO rise above 2000ppm the analyser’s pump stops
and its CO purge pump starts.
Your analyser displays P-OFF until CO level falls below 2000ppm.
Fuel type can be changed via “Menu”
or STATUS setting
CO/CO2 ratio
Pump Off, purge
Pump Off, purge
Excess air %
Pressure reading