1. 

BEFORE USING ANALYSER CHECK THE FOLLOWING:  

 

Particle filter is not dirty inside 
Water trap and flue probe hose are empty of water 
Water trap and red screw plug are fitted correctly to the analyser 
Flue probe hose is connected properly to the flue gas inlet 
Flue probe temperature plug is connected into T1 temperature connection 

 

Please read the Safety Warnings in the User Manual 

 
 

2. 

FRESH AIR PURGE 

 

Position the flue probe in fresh air, then press the “On/Off” button.  The analyser auto-
calibrates for approximately 90 seconds.  When complete… 
 
Select “Ratio” on the dial.  

In fresh air the CO reading = 0ppm

 

Select “O

2

/Eff” on the dial.  

In fresh air the O

2

 reading = 20.9% 

Select “Status” on the dial to view the following… 

 

STATUS display 

 

 

BAT         59 

11 : 46 : 29 
15 / 05 / 08 

CAL       283 

Battery status.  If less than 20 recharge or replace, (see section 10)

 

Current time.  Can be set via the “Menu”, (see section 11)

 

Current date.  Can be set via the “Menu”, (see section 11)

 

Number of days until next check and calibration is required

 

 
Note: Boiler inlet air temperature can either be… 
 

a) Set automatically by the flue probe during the fresh air purge 

   or 

b) Continuously measured if a thermocouple is plugged into the T2 socket 

 
 

3. 

COMBUSTION TESTS 

 

Select “Ratio” on the dial to check that the analyser is set for the correct fuel.  To change 
fuel select MENU / SETUP / SET FUEL then use scroll and enter, (see section 11). 
 
Position the flue probe as per the boiler manufacturer’s instructions; typically the tip of 
the flue probe is inserted to the centre of the flue.  The readings will stabilise after 60 
seconds assuming the boiler conditions are stable. 
 
The rotary switch can be used to display the following information… 

 
RATIO display 

 

 

NAT  GAS 

R        0.0001 
CO

P

          12 

CO

2

%     8.8 

Fuel type can be changed via “Menu”, (see section 11)

 

CO/CO

2

 ratio

 

Carbon Monoxide, (ppm)

 

Carbon Dioxide, (%)

 

 

Press SEND to print a full combustion test.  (Also sends to PC if Wireless module fitted). 
Hold SEND for 2+ seconds to log a full combustion report. 

Summary of Contents for 455

Page 1: ...rging indicator Analyser Connections Flue Gas Inlet see diagram overleaf see diagram overleaf ON OFF Turns the analyser ON OFF BACKLIGHT Switches backlight torch light ON OFF PUMP Turns the pump ON OF...

Page 2: ...connection Flue Gas Inlet Watertrap Drain with Red Screw Plug Flue probe temperature T1 Inlet temperature T2 Pressure connection P1 Pressure connection P2 differential Fit Spigot to Hose before connec...

Page 3: ...Current date Can be set via the Menu see section 11 Number of days until next check and calibration is required Note Boiler inlet air temperature can either be a Set automatically by the flue probe d...

Page 4: ...e pressure sensor Using the black connectors and manometer hose connect to P1 for single pressure or P1 and P2 for differential pressure PRS display PRESSURE P 0 04 mbar 12 56 29 Defaults to smoothing...

Page 5: ...be printed immediately or later from the memory Note The analyser s memory can store up to 20 tightness tests Tightness tests are logged automatically therefore the tightness section of the memory wil...

Page 6: ...HER DISPLAY CODES PO Pump Off OC Open Circuit on temperature input 9 FOR BEST PRESSURE SENSOR ACCURACY 1 Switch the analyser on for 5 minutes to let the temperature stabilise 2 Zero the pressure senso...

Page 7: ...minute TIGHTN S Set duration of tightness test in minutes Default 2 minute REPORT COMB N Stored combustion tests VIEW DEL ALL or EXIT max 99 tests PRESSURE Stored pressure tests VIEW DEL ALL or EXIT...

Page 8: ...o C 17 2 Appliance Customer NETT o C 37 9 Ref Appliance EFF C 98 3 LOSSES 1 7 Ref XAIR 34 8 C a l d u e o n 0 1 0 7 1 5 KANE455 V 1 00B YOUR COMPANY NAME PHONE NUMBER HERE SERIAL NO 000000000 CO CO2...

Page 9: ...tle Company Name Company Address 1 Address 2 Town City County Postcode Country Phone Number Fax Number Mobile Number Email Address Product Details Note Proof of Purchase may be required for warranty c...

Page 10: ...ecommendation Internet Search Exhibition Other Which do you read most often Often Sometimes Hardly Ever Registered Gas Engineer Gas Installer P H P I P H A M News Heating Ventilating Plumbing Heating...

Page 11: ......

Page 12: ...ying this analyser Before use please register on our website www kane co uk Scan the QR code to go directly to Register your Product on line or complete detach and return the Product Registration form...

Reviews: