COLD WEATHER PRECAUTIONS
Please keep your flue gas analyser in a warm place overnight.
Electronic devices become really cold when left in a vehicle overnight
then suffer when taken into a warm room. Condensation may form
affecting analyser performance & causing permanent damage.
Flue gas analysers are affected by condensation or water being sucked
into the analyser as their sensors can become blocked with water,
stopping them see flue gases. When this happens, Oxygen or Carbon
Dioxide reading will display as “—” & sensors may be permanently
damaged.
If you think your analyser is affected by condensation or water ingress,
you may be able to rectify the problem yourself. Leave your analyser
running in a warm place, with the pump ‘ON’ sampling fresh
air for a few hours whilst connected to mains power. If, after this, you
still experience problems please contact a KANE Service Centre or
authorised Service Partner.