4. MEASURING 

 
Make sure you do not exceed the meter’s operating specifications. 
 
Do not exceed the meter’s internal temperature operating range 
Do not put the meter on a hot surface 
 
When taking critical draft measurements always re-zero the meter in the position you 
are taking the readings and hold the meter still during the test. 
 
                                          

Differential Port          Port 1 

 

 

IR Printer Output 

 

Always use the top right hand port (Port 1) for taking single channel 
measurements (pressure or vacuum).  

Only use the left hand port for differential 

measurements. 

 

If the pressure being measured exceeds the meter’s design range the display will 
show ----

 

for “over-range”.

 

 
When taking draft readings at very low pressure or draft levels, for maximum 
accuracy, re-zero the meter in the orientation that it is being used.  This eliminated 
gravity effects on the very sensitive pressure transducer.  It is also recommended 
that the meter is switched on for at least five minutes and then re-zeroed before 
taking such sensitive measurements. 
 

Summary of Contents for 3200

Page 1: ...KANE3200 Differential Pressure Meter Stock No 18950 November 2011 Kane International Ltd...

Page 2: ...seconds until 3200 appears on the display During this 2 seconds all the segments of the display will be actuated The display will then show the following sequence 1 0 the firmware version HPA the unit...

Page 3: ...GE UNITS Quickly press and release The display will show the units currently set eg HPA Now press to scroll through the selection of units available When the units of choice are displayed press to sel...

Page 4: ...its are the ones required SAFETY WARNING Never connect to a pressure source where you are not sure what the maximum pressure might be Always ensuer that the meter you are using is correctly rated for...

Page 5: ...t hand port Port 1 for taking single channel measurements pressure or vacuum Only use the left hand port for differential measurements If the pressure being measured exceeds the meter s design range t...

Page 6: ...inus 200 mbar The meter does not auto range and so when Pa units are selected the full operating range cannot be displayed and the user must select an alternative range for measurements above 999 Pa A...

Page 7: ...ptable If not acceptable adjust the meter s position to minimize interference or switch off if possible the offending equipment during your test At the time of writing this manual November 2011 Kane I...

Page 8: ...Thank you for buying this Differential Pressure Meter Before use please register on our website www kane co uk...

Reviews: