KANE457 manual
Page
8
USING THE FOUR FUNCTION BUTTONS:
Switching ON the Analyser:
Press
/
button to switch the unit ON. This must be done in fresh air to
ensure that the analyser auto calibrates its’ sensors properly.
When switched on, the analyser beeps and briefly displays battery %, fuel and
pressure units. Its’ bottom line counts down from 90 until the sensors are ready to
use. If the analyser will not auto calibrate, its’ sensors need to be replaced or
recalibrated by a Kane authorised repair centre.
If an inlet temperature probe (optional) is connected into the T2 socket during its’
countdown, the measured temperature from the inlet probe will be used as the inlet
temperature.
If an inlet temperature probe is not connected to the analyser during countdown the
measured temperature from the flue probe will be used as the inlet temperature.
If neither probe is connected during countdown the analyser’s internal ambient
temperature will be used as the inlet temperature.
Switching OFF the Analyser
Press
/
button to switch the analyser OFF. The display counts down
from 30 with the pump on to clear the sensors with fresh air – If the probe is still
connected, make sure analyser and probe are in fresh air.
Press
/
if you want to stop the countdown and return to making
measurements.
Note: The analyser will not switch off unless the CO reading is below 20ppm.
Backlight & Tasklight
Press
/
to switch the display's backlight and tasklight on and off.
NOTE: Use of the backlight/tasklight significantly increases the current drain on
the batteries.
Switching PUMP on / off
The analyser normally operates with the pump on.
Press
/
to switch the pump off and on.