Kane 100-1 Manual Download Page 3

Page 

3

 

 

English 18259 

 

1.    BEFORE USING THE ANALYSER FOR THE FIRST TIME: 

 

The analyser may be supplied with either or both of the following chargers. 
To fully charge the battery, you may use a Kane trickle charger, plugged into the base 
of the unit for 12-24 hours. Alternatively the Kane fast charger which plugs into the 
side  will take about 1-2 hours, after which it will switch to low trickle charge. Do not 
fast charge the unit unless the battery charge indicator reads less than 70% otherwise 
over heating could cause an internal thermal cutout to operate. The unit will then not 
operate on its internal battery until it has cooled.  
Avoid leaving the unit on charge and switched off for more than 1 –2 days at a time 
or the internal battery life could be reduced.  
 
Set the correct time, date etc, after it is switched on and calibrated – See 

USING 

THE ROTARY DIAL

 below. These are stored when the analyser is switched off. 

 
 

2.  BEFORE USING THE ANALYSER EVERY TIME: 

 

Check the water trap is empty and the particle filter is not dirty:  
 

To empty the water trap, unplug its rubber stopper and re-plug once it is empty.  

 

To change the filter, remove protective rubber sleeve, pull out the water trap , 
remove the water trap’s particle filter from the  spigot and replace. Reconnect 
the water trap and rubber protective sleeve.  

 
After switch on check that date and time are correct and battery power is sufficient.  
 
Avoid exposing the instrument to sudden large temperature changes, and ensure that 
the unit reads zero for CO and approx. 400 ppm CO

2

 in fresh air before use. If 

necessary re-zero (see: Menu Selection). 
 
Avoid very close proximity to radio transmitting devices, otherwise readings may be 
affected. 

 

 

SAFETY WARNING 

 
This analyser can be used to extract combustion gases that may be toxic in relatively 
low concentrations. These gases are exhausted from the back of the instrument. 

This 

analyser must only be used in well-ventilated locations by trained and competent 
persons after due consideration of all the potential hazards.

 

 

Summary of Contents for 100-1

Page 1: ...onoxide and Ambient Carbon Dioxide Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Tel 44 0 1707 375550 Fax 44 0 1707 393277 E mail sales kane co uk www kane co...

Page 2: ...when in menu mode The rotary dial changes the display s 2nd line and selects access to the menu to make changes to the date time fuel etc The KANE 100 1 can store up to 255 sets of readings The Kane 1...

Page 3: ...witched off 2 BEFORE USING THE ANALYSER EVERY TIME Check the water trap is empty and the particle filter is not dirty To empty the water trap unplug its rubber stopper and re plug once it is empty To...

Page 4: ...0ppm in fresh air allow a few seconds after each button press for readings to settle and then press ENTER to confirm After ENTER is pressed it will take about 4 seconds before the unit is ready to res...

Page 5: ...keys to change settings such as time and date when the rotary dial is turned to MENU Enter key to Hold or Store readings In normal operation not MENU the ENTER key is used to either HOLD the readings...

Page 6: ...on top of the analyser Switching PUMP on off The analyser normally operates with the pump on Press the PUMP key quickly to switch the pump off and on When the pump is switched off the analyser displa...

Page 7: ...ribed fully below to select and change the following functions 1 Time Uses Military time as standard 7am 07 00 7pm 19 00 2 Date with date format selection 3 Pump speed control full speed or quiet mode...

Page 8: ...final logical ENTER is pressed no changes are made Note in AIR ZERO this mode will continue until terminated by the pressing of ENTER You can choose from the menu fresh air zeroing which allows CO2 t...

Page 9: ...ber within the range of data stored up to 255 maximum In view mode only stored readings wrap round so that using the up key you pass from maximum reading back to the first reading or using the down ke...

Page 10: ...You can maximize the number of stored readings by switching the pump to low speed QUIET and after initiating store switch the pump off Typically if the batteries are fully charged you will be able to...

Page 11: ...fresh air zero or true zero Use the UP or Down button to chose and ENTER to select Once zeroing has commenced it will continue until completed whatever you do with the rotary dial Fresh air zero shoul...

Page 12: ...operating specifications In particular Do not exceed the probe s maximum temperature any plastic parts must not contact hot flues Do not exceed the temperature operating range Do not put the analyser...

Page 13: ...ndard printouts are Instant print press print button KANE 100 1 CO CO2 MONITOR DATE 01 01 06 TIME 14 27 08 CO2 CO ppm ppm 950 12 Manually logged print via VIEW stored data KANE 100 1 CO CO2 MONITOR DA...

Page 14: ...via MEMORY KANE 100 1 CO CO2 MONITOR AUTOLOG REPORT DATE 01 01 06 TIME 18 27 08 Start Time 06 6 04 14 00 00 Stop Time 06 6 04 17 59 00 No of readings 255 Max CO2 Av CO2 Max CO Av CO ppm ppm ppm ppm 95...

Page 15: ...start up 8 PROBLEM SOLVING If any problems are not solved with these solutions contact us or an authorized repair centre Fault symptom Causes Solutions Batteries not holding charge Batteries exhausted...

Page 16: ...NU appears on the top line NB You may be changing settings when you do this Rotate the dial to the required display position The unit will not turn off Do not hold down the on off button Try connectin...

Page 17: ...4000 ppm 2000ppm 9999ppm Dimensions Weight Handset 1kg 2 2lb 200mm 7 9 x 45mm 1 8 x 90mm 3 5 Ambient Operating Range 0 C to 45 C 32 104 F 10 to 90 RH non condensing Battery Life 6 hours from full char...

Page 18: ...tart up sequence in the location where the analyser will be used Switch on all localized electrical equipment capable of causing interference Check all readings are as expected A level of disturbance...

Page 19: ...mg m3 8 hours 9ppm 10mg m3 Carbon Dioxide HSE Regulation EH40 sets the maximum occupational exposure limit at 5000ppm BSRIA recommend a maximum of 800ppm over an eight hour time weighted average ASHRA...

Reviews: