English 18259
AUTOSTORE
Hold will be inhibited during auto store. If you push the Hold button during auto
store AUTO STO will be indicated on the top line. It is recommended that you select
ZERO via the menu and perform a zero before starting a series of Auto stored
readings.
Memory is always cleared before a new autostore sequence is initiated. Be sure to
print out your previous results before starting a new sequence.
When store intervals greater than one minute are set and the pump is switched off it
will power up automatically one minute before a reading is due to be stored. The
pump will run whatever speed (high or low ) was previously set. Where the store
interval is set to 1 minute the pump momentarily switches off every minute.
You can maximize the number of stored readings by switching the pump to low
speed (QUIET) and after initiating store switch the pump off. Typically if the
batteries are fully charged you will be able to store 255 readings at up to 3 minutes
interval.
To store for longer periods above 12 hours you should connect the Kane trickle
charger.
The top line displays AUTO STO
Use the UP/DOWN keys to select from OFF or ON
Press ENTER to select
The display will show:
AUTO
STO
TIME
01
Use the UP /DOWN keys to set any store interval
between 1 and 10 minutes.
The display top line will then revert to MENU
Select any required position on the rotary dial.
Every time readings are stored the top line will flash LOG and the STORE number
until the memory is full or until the autostore is switched off. There will also be a
beep for every store to a maximum of 255 sets of readings.
You can only exit AUTOSTORE by returning to MENU mode and setting
AUTOSTO to OFF.