background image

4

below your desired comparison temperature, say +4°C or +7°C as 
appropriate. You can use your reference thermometer for this.

(0.5 litres of water in a plastic beaker at +7°C with an ambient 
temperature of 20°C exhibits a temperature rise of less than 0.1°C per 
minute.)

Switch on both thermometers and insert both the probes into the water 
to a depth of at least 50mm (2 inches). Hold the probes so that their tips 
are very close together and use the probes to stir the water. The water 
should be stirred in the centre of the beaker away from the sides. The 
agitation rate should be approximately one rotation per second.

Once the KM20REF is displaying the required temperature of 
comparison and this reading has been stable for 30 secs the reading of 
the food thermometer can be taken and recorded.

If the reading from the food thermometer differs from the traceable 
reading of the KM20REF by ±0.5°C or more, the food thermometer and 
probe should be returned to the manufacturer for recalibration.

Note: 

 The traceable KM20REF reading is obtained by adding the 

appropriate adjustment. This adjustment is obtained by taking the 
applied temperature and subtracting the corresponding measured 
value. See the example below:

Applied Temperature 

Measured 

Adjustment

 –18.0°C 

–18.2°C 

+0.2

 0.0°C 

–0.1°C  +0.1

 +5.0°C 

+5.1°C 

–0.1

 +8.0°C 

+8.1°C 

–0.1

 +20.0°C 

+20.2°C 

–0.2

Note:

  In an ideal world your tests should be conducted at exactly 

the same points as those indicated in the “Measured” column of your 
KM20REF Certifi cate. For all practical purposes it is acceptable to 
apply the adjustment factor provided the readings are ±1.0°C of the 
“Measured” value.

If then when conducting the test the following results were achieved 

KM20REF reading 

+8.1°C

Food thermometer reading 

+8.5°C

your results sheet would be fi lled in as follows:

Summary of Contents for KM20REF

Page 1: ...KM20REF REFERENCE THERMOMETER...

Page 2: ...Support department or your local distributor Repeated sharp flexing can break thermocouple probe leads To prolong lead life avoid sharp bends or kinks in the leads especially near the connector This...

Page 3: ...one of the finest in the UK Comark UKAS certificates can offer the lowest uncertainty of 0 01 C and provide independent proof of correct calibration using equipment and procedures audited by UKAS ins...

Page 4: ...line battery is used and that this is replaced after 25 hours of use Food temperature measuring systems comprise two parts the hand held meter and the sensor probe It is important that the accuracy of...

Page 5: ...f the reading from the food thermometer differs from the traceable reading of the KM20REF by 0 5 C or more the food thermometer and probe should be returned to the manufacturer for recalibration Note...

Page 6: ...KM20REF SERIAL NUMBER 1234 FOOD THERMOMETER SERIAL NUMBERS INSTRUMENT 0675 PROBE 0231 KM20REF READING 8 1 C ADJUSTMENT 0 1 C TRACEABLE TEMPERATURE 8 0 C FOOD THERMOMETER READING 8 5 C DEVIATION 0 5 C...

Page 7: ...TEMPERATURE C FOOD THERMOMETER READING C DEVIATION C ACCEPT REJECT TESTED BY CHECKED BY Auto Switch Off The KM20REF has an auto switch off function to conserve battery life The instrument will automat...

Page 8: ...EASUREMENT RANGE Pt100 100 C to 199 9 C RESOLUTION 0 1 C ACCURACY 0 2 C OPERATING RANGE 0 to 40 C CERTIFICATE OF CALIBRATION DESIGNATED POINTS FOR ACCURACY 18 C 0 C 5 C 8 C 20 C EMC Tested to EN 61326...

Page 9: ...8...

Page 10: ...Norwich Norfolk NR6 6JB United Kingdom Tel 44 844 815 6599 Fax 44 844 815 6598 Email sales comarkinstruments com Website www comarkinstruments com Comark Instruments PO Box 500 Beaverton OR 97077 USA...

Reviews: